![Tiger Optics HALO M7000 Series Instruction Manual Download Page 14](http://html.mh-extra.com/html/tiger-optics/halo-m7000-series/halo-m7000-series_instruction-manual_1115521014.webp)
HALO Users Manual
Revision G
6
Table 2-2 HALO-CH
4
Trace Gas Analyzer Specifications
Performance
Lowest Detection Limit
4 ppb
Sensitivity
2 ppb
Accuracy
4% of reading or +/- 2 ppb if greater
Speed of Response
(typical)
Dry Down Response to 50ppb intrusion - 90% of final value in
<1min
Operating Range
0 – 8000 ppb
Environmental Conditions 10º - 40ºC - - - 30% to 80% R.H. Non-condensing
Storage Temperature
-10°C to 50°C
Technology
Method
Cavity Ring-Down Spectroscopy (CRDS)
Laser Light Source
Continuous Wave (CW) Near-IR diode laser
Approvals
CE Marked – LVD and EMC approved
Patents
U.S. Patent # 5,528,040
U.S. Patents Pending
Gas Sample Conditions
Sample Inlet Pressure
30 – 125 psig (3.1 – 9.6 bara)
Sample Outlet Pressure
< 2.0 psig back pressure (or vacuum)
Flow Rate
0.5 to 1.8 slpm depending on gas type
Sample Gases
N
2
, He, Ar, H
2
, and O
2
(contact Tiger Optics for additions to the
list of sample gases).
Sample line temperature
Up to 60ºC