33
10. Technical data
High-voltage Insulation Resistance Tester
ΩMegaSafe ISO 1Ex in accordance with
EN/IEC 61010-1 and EN/IEC 61557-2 (VDE 0413 part 2)
Explosionproof type in accordance with EN 60079-0
and EN 60079-11
EC-type examination certificate (ISO 1 Ex)
II 2 G Ex ib IIBT4 BVS 06 ATEX E130
I M2 (M1) EX ib [ia] I BVS 06 ATEX E130
Safety-related characteristics:
max. output voltage 550 V
max. output current 1,5 mA
max. inductance 28 H
max. capacitance 1 nF / 500 V
100 nF / 100 V
200 nF / 32 V
Measurement range:
1,999 MΩ
Resolution 1 kΩ 5% +/- 5D
19,99 MΩ
Resolution 10 kΩ 5% +/- 5D
1,999 GΩ
Resolution 1 MΩ 5% +/- 5D
Additionally with selected measurement option TΩ:
19,99 GΩ
Resolution 10 MΩ 10% +/- 5D
199,9 GΩ
Resolution 100 MΩ 10% +/- 5D
1999 GΩ
Resolution 1 GΩ 25% +/- 5D
Test voltage/measuring current:
32 V 64 µA
+/-20 %
100 V 200 µA
+/-20 %
500 V 0,5 mA
+/-20 %
Discharge:
for rapid discharge of
capacitive voltage
Digital display:
doublespaced LCD display,
3 1/2-digit, overflow display
through OL screen backlighting
Voltage indication:
red LED and acoustic warning
signal and indication of the value
in Volt
24 - 400 V AC
15 - 500 V DC
Surge voltage category:
CAT III 600 V in accordance with
EN/IEC 61010-1
Operating temperature:
-10 ... +50°C
Power supply:
battery 9 V block AlMn IEC 6LR61
Duracell MN 1604 Plus Power
Duralock (6LF22)
(type, see instructions)
multi-stage BAT indication
automatic switch-off
Casing:
made of impact-proof ABS with
unbreakable display cover
antistatic coating
Voltage surge protection:
IP 65,
usable in precipitatio
n
Weight/Dimensions:
60 x 230 x 40 mm / 220 g