29
5.5
Volume resistance wirh electrode for surface
resistance and volume resistivity for testing
on surfaces, art.-no. 73012:
e. g. insulation materials or conductive rubber pads
Connect the tripod electrode and the upper
socket of the ΩMegaSafe ISO 1Ex with the
measuring line
Place the test object on a good conductive
underground.
Connect the the conductive underground with
the lower socket of the ΩMegaSafe ISO 1Ex.
Place the tripod electrodes on the test object
and weight it with approx. 30 kg.
Test the volume resistance with the
ΩMegaSafe ISO 1Ex.
5.6
Volumene rsistance with tripod electrode:
e.g. large-area materials
Connect the electrode for surface resistance and
volume resistivity and the upper socket of the
ΩMegaSafe ISO 1Ex with the measuring line
Connect the conductive underground with the
lower socket of the ΩMegaSafe ISO 1Ex.
Place the test object on the conductive under
ground and place the test electrode on the test
object
Test the volume resistance with the ΩMegaSafe
ISO 1Ex.
test object
electrode at once
standard weight
2,5 kg
conductive
underground
conductive
underground
test object
tripod electrode
30 kg