24
VSP63-de-120207
4.2.2
Adjustment On Zero Reference Pressure
For this adjustment a pressure below 1.0x10
-2
mbar has to be applied.
To achieve optimum results of the adjustment we recommend
a warm-up of at least 5 minutes at the appropriate calibration
pressure before any adjustment. Important: be sure to use
suitable reference gauges only!
For adjustment first remove the rubber caps from the "adj" and "set" button (1),
then push the "set" button by means of a screw driver or other suitable tool (2).
Now a voltage is provided at the signal output which corresponds to the actual
zero reference pressure (ex works: 1,4V for "ur" pressure underrange).
By pushing the "adj" button several times you can now increment this value to
maximal 1.0x10
-2
mbar (3) according to the pressure-voltage relation in
chapter 4.1. Then the output again starts at the voltage for "ur". When the
voltage output is correctly adjusted to the actual reference pressure, the
adjustment procedure is carried out automatically after 5s. Further the new
reference pressure for zero adjustment is saved. From now on zero adjustment
as described in chapter 4.2.1 (push "adj" button at zero pressure) will result in an
adjustment on the saved reference pressure. Insert the rubber caps again.
When the "set" button is pushed without further action you can check the
currently saved zero reference pressure. The transducer then outputs the
corresponding voltage for 5s, no adjustment is carried out.
Summary of Contents for VSP63
Page 1: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 14: ...14 VSP63 de 120207 Konformit tserkl rung...
Page 27: ...27 VSP63 de 120207 Declaration of Conformity...
Page 28: ...VSP63 de 120207...
Page 29: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 42: ...14 VSP63 de 120207 Konformit tserkl rung...
Page 55: ...27 VSP63 de 120207 Declaration of Conformity...
Page 56: ...VSP63 de 120207...
Page 57: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 70: ...14 VSP63 de 120207 Konformit tserkl rung...
Page 83: ...27 VSP63 de 120207 Declaration of Conformity...
Page 84: ...VSP63 de 120207...
Page 85: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 98: ...14 VSP63 de 120207 Konformit tserkl rung...
Page 111: ...27 VSP63 de 120207 Declaration of Conformity...
Page 112: ...VSP63 de 120207...
Page 113: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 126: ...14 VSP63 de 120207 Konformit tserkl rung...
Page 139: ...27 VSP63 de 120207 Declaration of Conformity...
Page 140: ...VSP63 de 120207...