20 

VSP63-de-120207

 

 
 

 
When mounting the transducer avoid forced twisting or violent 
opening. The housing can be mechanically damaged this way!  

 
 

 

 

Overpressure in the vacuum system > 1 bar 
Accidental or unintended opening of clamp elements under 
stress can lead to injuries due to parts flying around! 
Unsecured hose connections can release, process media thus 
can leak and possibly damage your health. 

 

 
 

Overpressure in the vacuum system 1,5 to 4 bar 
KF flange connections with elastomer sealings cannot withstand 
such pressures.  Process media thus can leak and possibly 
damage your health. 
 
 
 

 

3.3 

Electrical Connection 

3.3.1 

Operation With Thyracont Display Unit 

For operation of the transducer with a Thyracont display and control unit a 
suitable measurement cable must be used (see accessories).  
 
 

 
Do not connect or disconnect the transducer when the cable is 
on circuit!  

 
 

 
 
Connect the cables plug to the transducer and secure it with the screw.  Connect 
the other end of the cable to the display unit and secure the plug. Only now 
connect your display unit to mains power or switch it on respectively.  
 
 
 
 

Summary of Contents for VSP63

Page 1: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...

Page 2: ...Thyracont Anzeigeger te 7 3 3 2 Kundeneigene Spannungsversorgung 8 4 Betrieb 9 4 1 Allgemeines 9 4 2 Nachjustieren 10 4 2 1 Abgleich auf Atmosph re und Nullpunkt 10 4 2 2 Abgleich auf einen bestimmten...

Page 3: ...gliche Reaktionen zwischen Werkstoffen und Prozessmedien Ber cksichtigen Sie m gliche Reaktionen der Prozessmedien infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver nd...

Page 4: ...zum Anschluss an Anzeigeger t VD9 W0606006 Gegenstecker 6polig gewinkelt XB0600002 2 3 Produktbeschreibung Der Vakuum Messumformer VSP63 dient zur Absolutdruck Messung im Bereich 1000 1 0x10 4 mbar Da...

Page 5: ...rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt der Einsatz zu Zwecken die von oben genannten abweichen insbesondere der Anschluss an Ger te oder Komponenten die laut ihr...

Page 6: ...leinflansch DN16 ISO KF VSP63MV VSP63MVSW oder DN16CF Flansch VSP63MVCF herstellen Bei Verwendung des Schlauchanschlusses von Modell VSP63MVSW ist die Schlauchverbindung geeignet zu sichern z B mittel...

Page 7: ...ei KF Flanschverbindungen k nnen Elastomer Dichtringe dem Druck nicht mehr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschlus...

Page 8: ...EMV gerecht gem untenstehender Pinbelegung herzustellen Pinbelegung 1 2 3 6 5 4 Stecker Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Typ Hirschmann 6pol m nnl Leitungs Abschluss LA 3 0 kOhm auf GND Signalausga...

Page 9: ...insbesondere nach extremen Druckspr ngen angebracht sein eine Stabilisierungszeit von 2min zu beachten Messgenauigkeit Das Ger t ist ab Werk in stehender Position bei einer Versorgungsspannung von 24V...

Page 10: ...m welchen Justierpunkt es sich handelt Beim Nullabgleich sollte der Ist Druck kleiner 5 0x10 5 mbar sein alternativ kann auf einen einstellbaren Referenzwert abgeglichen werden siehe unten Um optimale...

Page 11: ...Nullpunkt Justierwert entspricht Werkseinstellung 1 4V f r ur underrange Durch mehrmaliges Dr cken der Taste adj kann dieser Wert nun gem des Zusammenhangs zwischen Druck und Ausgangssignal aus Abschn...

Page 12: ...nden Sie das Ger t bitte mit einer Kontaminationserkl rung zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Wendel...

Page 13: ...genauigkeit 20 0 001mbar 2 v Messwert Reaktionszeit 200 ms Spannungsversorgung 15 30 VDC Elektrischer Anschluss Typ Hirschmann 6polig m nnl verschraubbar Leistungsaufnahme max 1 0 W bei 24 V Versorgun...

Page 14: ...14 VSP63 de 120207 Konformit tserkl rung...

Page 15: ...Thyracont Display Unit 20 3 3 2 Operation With Other Supply And Evaluation Units 21 4 Operation 22 4 1 General 22 4 2 Adjustment 23 4 2 1 Adjustment On Atmosphere and Zero 23 4 2 2 Adjustment On Zero...

Page 16: ...possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversions or modif...

Page 17: ...2m for connection to VD9 display unit W0606002 Measurement cable 6m for connection to VD9 display unit W0606006 Counterplug 6pole 90 bent XB0600002 2 3 Product Description The VSP63 vacuum transducer...

Page 18: ...class A and therefore can cause radio interference in living quarters Improper Use The use for purposes not covered above is regarded as improper in particular the connection to components not allowe...

Page 19: ...again during maintenance work Make vacuum connection via small flange DN16 ISO KF VSP63MV VSP63MVSW or DN16CF Flansch VSP63MVCF If you use the hose barb of model VSP63MVSW be sure to secure the hose...

Page 20: ...F flange connections with elastomer sealings cannot withstand such pressures Process media thus can leak and possibly damage your health 3 3 Electrical Connection 3 3 1 Operation With Thyracont Displa...

Page 21: ...considering EMI demands and according to the pin description shown below Pin Description 1 2 3 6 5 4 Socket Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Type Hirschmann 6pole male Line Terminating Resistor 3...

Page 22: ...y of the unit it is appropriate to allow for stabilization time of 2 5 minutes especially when extreme pressure changes have occurred Accuracy The unit is adjusted ex works in upright position and at...

Page 23: ...ch adjustment point is relevant For zero adjustment actual pressure should be less than 5 0x10 5 mbar alternatively the unit can be adjusted on a variable zero pressure reference see below To achieve...

Page 24: ...ressure underrange By pushing the adj button several times you can now increment this value to maximal 1 0x10 2 mbar 3 according to the pressure voltage relation in chapter 4 1 Then the output again s...

Page 25: ...P63 please send the instrument to us for repair and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or break...

Page 26: ...or 2 from reading Repeatability 20 0 001mbar 2 from reading Response Time 200 ms Power Supply 15 30 VDC Electrical Connection Type Hirschmann 6pin male lockable Power Consumption max 1 0 W with 24 V s...

Page 27: ...27 VSP63 de 120207 Declaration of Conformity...

Page 28: ...VSP63 de 120207...

Page 29: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...

Page 30: ...Thyracont Anzeigeger te 7 3 3 2 Kundeneigene Spannungsversorgung 8 4 Betrieb 9 4 1 Allgemeines 9 4 2 Nachjustieren 10 4 2 1 Abgleich auf Atmosph re und Nullpunkt 10 4 2 2 Abgleich auf einen bestimmten...

Page 31: ...gliche Reaktionen zwischen Werkstoffen und Prozessmedien Ber cksichtigen Sie m gliche Reaktionen der Prozessmedien infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver nd...

Page 32: ...zum Anschluss an Anzeigeger t VD9 W0606006 Gegenstecker 6polig gewinkelt XB0600002 2 3 Produktbeschreibung Der Vakuum Messumformer VSP63 dient zur Absolutdruck Messung im Bereich 1000 1 0x10 4 mbar Da...

Page 33: ...rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt der Einsatz zu Zwecken die von oben genannten abweichen insbesondere der Anschluss an Ger te oder Komponenten die laut ihr...

Page 34: ...leinflansch DN16 ISO KF VSP63MV VSP63MVSW oder DN16CF Flansch VSP63MVCF herstellen Bei Verwendung des Schlauchanschlusses von Modell VSP63MVSW ist die Schlauchverbindung geeignet zu sichern z B mittel...

Page 35: ...ei KF Flanschverbindungen k nnen Elastomer Dichtringe dem Druck nicht mehr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschlus...

Page 36: ...EMV gerecht gem untenstehender Pinbelegung herzustellen Pinbelegung 1 2 3 6 5 4 Stecker Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Typ Hirschmann 6pol m nnl Leitungs Abschluss LA 3 0 kOhm auf GND Signalausga...

Page 37: ...insbesondere nach extremen Druckspr ngen angebracht sein eine Stabilisierungszeit von 2min zu beachten Messgenauigkeit Das Ger t ist ab Werk in stehender Position bei einer Versorgungsspannung von 24V...

Page 38: ...m welchen Justierpunkt es sich handelt Beim Nullabgleich sollte der Ist Druck kleiner 5 0x10 5 mbar sein alternativ kann auf einen einstellbaren Referenzwert abgeglichen werden siehe unten Um optimale...

Page 39: ...Nullpunkt Justierwert entspricht Werkseinstellung 1 4V f r ur underrange Durch mehrmaliges Dr cken der Taste adj kann dieser Wert nun gem des Zusammenhangs zwischen Druck und Ausgangssignal aus Abschn...

Page 40: ...nden Sie das Ger t bitte mit einer Kontaminationserkl rung zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Wendel...

Page 41: ...genauigkeit 20 0 001mbar 2 v Messwert Reaktionszeit 200 ms Spannungsversorgung 15 30 VDC Elektrischer Anschluss Typ Hirschmann 6polig m nnl verschraubbar Leistungsaufnahme max 1 0 W bei 24 V Versorgun...

Page 42: ...14 VSP63 de 120207 Konformit tserkl rung...

Page 43: ...Thyracont Display Unit 20 3 3 2 Operation With Other Supply And Evaluation Units 21 4 Operation 22 4 1 General 22 4 2 Adjustment 23 4 2 1 Adjustment On Atmosphere and Zero 23 4 2 2 Adjustment On Zero...

Page 44: ...possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversions or modif...

Page 45: ...2m for connection to VD9 display unit W0606002 Measurement cable 6m for connection to VD9 display unit W0606006 Counterplug 6pole 90 bent XB0600002 2 3 Product Description The VSP63 vacuum transducer...

Page 46: ...class A and therefore can cause radio interference in living quarters Improper Use The use for purposes not covered above is regarded as improper in particular the connection to components not allowe...

Page 47: ...again during maintenance work Make vacuum connection via small flange DN16 ISO KF VSP63MV VSP63MVSW or DN16CF Flansch VSP63MVCF If you use the hose barb of model VSP63MVSW be sure to secure the hose...

Page 48: ...F flange connections with elastomer sealings cannot withstand such pressures Process media thus can leak and possibly damage your health 3 3 Electrical Connection 3 3 1 Operation With Thyracont Displa...

Page 49: ...considering EMI demands and according to the pin description shown below Pin Description 1 2 3 6 5 4 Socket Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Type Hirschmann 6pole male Line Terminating Resistor 3...

Page 50: ...y of the unit it is appropriate to allow for stabilization time of 2 5 minutes especially when extreme pressure changes have occurred Accuracy The unit is adjusted ex works in upright position and at...

Page 51: ...ch adjustment point is relevant For zero adjustment actual pressure should be less than 5 0x10 5 mbar alternatively the unit can be adjusted on a variable zero pressure reference see below To achieve...

Page 52: ...ressure underrange By pushing the adj button several times you can now increment this value to maximal 1 0x10 2 mbar 3 according to the pressure voltage relation in chapter 4 1 Then the output again s...

Page 53: ...P63 please send the instrument to us for repair and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or break...

Page 54: ...or 2 from reading Repeatability 20 0 001mbar 2 from reading Response Time 200 ms Power Supply 15 30 VDC Electrical Connection Type Hirschmann 6pin male lockable Power Consumption max 1 0 W with 24 V s...

Page 55: ...27 VSP63 de 120207 Declaration of Conformity...

Page 56: ...VSP63 de 120207...

Page 57: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...

Page 58: ...Thyracont Anzeigeger te 7 3 3 2 Kundeneigene Spannungsversorgung 8 4 Betrieb 9 4 1 Allgemeines 9 4 2 Nachjustieren 10 4 2 1 Abgleich auf Atmosph re und Nullpunkt 10 4 2 2 Abgleich auf einen bestimmten...

Page 59: ...gliche Reaktionen zwischen Werkstoffen und Prozessmedien Ber cksichtigen Sie m gliche Reaktionen der Prozessmedien infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver nd...

Page 60: ...zum Anschluss an Anzeigeger t VD9 W0606006 Gegenstecker 6polig gewinkelt XB0600002 2 3 Produktbeschreibung Der Vakuum Messumformer VSP63 dient zur Absolutdruck Messung im Bereich 1000 1 0x10 4 mbar Da...

Page 61: ...rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt der Einsatz zu Zwecken die von oben genannten abweichen insbesondere der Anschluss an Ger te oder Komponenten die laut ihr...

Page 62: ...leinflansch DN16 ISO KF VSP63MV VSP63MVSW oder DN16CF Flansch VSP63MVCF herstellen Bei Verwendung des Schlauchanschlusses von Modell VSP63MVSW ist die Schlauchverbindung geeignet zu sichern z B mittel...

Page 63: ...ei KF Flanschverbindungen k nnen Elastomer Dichtringe dem Druck nicht mehr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschlus...

Page 64: ...EMV gerecht gem untenstehender Pinbelegung herzustellen Pinbelegung 1 2 3 6 5 4 Stecker Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Typ Hirschmann 6pol m nnl Leitungs Abschluss LA 3 0 kOhm auf GND Signalausga...

Page 65: ...insbesondere nach extremen Druckspr ngen angebracht sein eine Stabilisierungszeit von 2min zu beachten Messgenauigkeit Das Ger t ist ab Werk in stehender Position bei einer Versorgungsspannung von 24V...

Page 66: ...m welchen Justierpunkt es sich handelt Beim Nullabgleich sollte der Ist Druck kleiner 5 0x10 5 mbar sein alternativ kann auf einen einstellbaren Referenzwert abgeglichen werden siehe unten Um optimale...

Page 67: ...Nullpunkt Justierwert entspricht Werkseinstellung 1 4V f r ur underrange Durch mehrmaliges Dr cken der Taste adj kann dieser Wert nun gem des Zusammenhangs zwischen Druck und Ausgangssignal aus Abschn...

Page 68: ...nden Sie das Ger t bitte mit einer Kontaminationserkl rung zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Wendel...

Page 69: ...genauigkeit 20 0 001mbar 2 v Messwert Reaktionszeit 200 ms Spannungsversorgung 15 30 VDC Elektrischer Anschluss Typ Hirschmann 6polig m nnl verschraubbar Leistungsaufnahme max 1 0 W bei 24 V Versorgun...

Page 70: ...14 VSP63 de 120207 Konformit tserkl rung...

Page 71: ...Thyracont Display Unit 20 3 3 2 Operation With Other Supply And Evaluation Units 21 4 Operation 22 4 1 General 22 4 2 Adjustment 23 4 2 1 Adjustment On Atmosphere and Zero 23 4 2 2 Adjustment On Zero...

Page 72: ...possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversions or modif...

Page 73: ...2m for connection to VD9 display unit W0606002 Measurement cable 6m for connection to VD9 display unit W0606006 Counterplug 6pole 90 bent XB0600002 2 3 Product Description The VSP63 vacuum transducer...

Page 74: ...class A and therefore can cause radio interference in living quarters Improper Use The use for purposes not covered above is regarded as improper in particular the connection to components not allowe...

Page 75: ...again during maintenance work Make vacuum connection via small flange DN16 ISO KF VSP63MV VSP63MVSW or DN16CF Flansch VSP63MVCF If you use the hose barb of model VSP63MVSW be sure to secure the hose...

Page 76: ...F flange connections with elastomer sealings cannot withstand such pressures Process media thus can leak and possibly damage your health 3 3 Electrical Connection 3 3 1 Operation With Thyracont Displa...

Page 77: ...considering EMI demands and according to the pin description shown below Pin Description 1 2 3 6 5 4 Socket Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Type Hirschmann 6pole male Line Terminating Resistor 3...

Page 78: ...y of the unit it is appropriate to allow for stabilization time of 2 5 minutes especially when extreme pressure changes have occurred Accuracy The unit is adjusted ex works in upright position and at...

Page 79: ...ch adjustment point is relevant For zero adjustment actual pressure should be less than 5 0x10 5 mbar alternatively the unit can be adjusted on a variable zero pressure reference see below To achieve...

Page 80: ...ressure underrange By pushing the adj button several times you can now increment this value to maximal 1 0x10 2 mbar 3 according to the pressure voltage relation in chapter 4 1 Then the output again s...

Page 81: ...P63 please send the instrument to us for repair and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or break...

Page 82: ...or 2 from reading Repeatability 20 0 001mbar 2 from reading Response Time 200 ms Power Supply 15 30 VDC Electrical Connection Type Hirschmann 6pin male lockable Power Consumption max 1 0 W with 24 V s...

Page 83: ...27 VSP63 de 120207 Declaration of Conformity...

Page 84: ...VSP63 de 120207...

Page 85: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...

Page 86: ...Thyracont Anzeigeger te 7 3 3 2 Kundeneigene Spannungsversorgung 8 4 Betrieb 9 4 1 Allgemeines 9 4 2 Nachjustieren 10 4 2 1 Abgleich auf Atmosph re und Nullpunkt 10 4 2 2 Abgleich auf einen bestimmten...

Page 87: ...gliche Reaktionen zwischen Werkstoffen und Prozessmedien Ber cksichtigen Sie m gliche Reaktionen der Prozessmedien infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver nd...

Page 88: ...zum Anschluss an Anzeigeger t VD9 W0606006 Gegenstecker 6polig gewinkelt XB0600002 2 3 Produktbeschreibung Der Vakuum Messumformer VSP63 dient zur Absolutdruck Messung im Bereich 1000 1 0x10 4 mbar Da...

Page 89: ...rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt der Einsatz zu Zwecken die von oben genannten abweichen insbesondere der Anschluss an Ger te oder Komponenten die laut ihr...

Page 90: ...leinflansch DN16 ISO KF VSP63MV VSP63MVSW oder DN16CF Flansch VSP63MVCF herstellen Bei Verwendung des Schlauchanschlusses von Modell VSP63MVSW ist die Schlauchverbindung geeignet zu sichern z B mittel...

Page 91: ...ei KF Flanschverbindungen k nnen Elastomer Dichtringe dem Druck nicht mehr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschlus...

Page 92: ...EMV gerecht gem untenstehender Pinbelegung herzustellen Pinbelegung 1 2 3 6 5 4 Stecker Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Typ Hirschmann 6pol m nnl Leitungs Abschluss LA 3 0 kOhm auf GND Signalausga...

Page 93: ...insbesondere nach extremen Druckspr ngen angebracht sein eine Stabilisierungszeit von 2min zu beachten Messgenauigkeit Das Ger t ist ab Werk in stehender Position bei einer Versorgungsspannung von 24V...

Page 94: ...m welchen Justierpunkt es sich handelt Beim Nullabgleich sollte der Ist Druck kleiner 5 0x10 5 mbar sein alternativ kann auf einen einstellbaren Referenzwert abgeglichen werden siehe unten Um optimale...

Page 95: ...Nullpunkt Justierwert entspricht Werkseinstellung 1 4V f r ur underrange Durch mehrmaliges Dr cken der Taste adj kann dieser Wert nun gem des Zusammenhangs zwischen Druck und Ausgangssignal aus Abschn...

Page 96: ...nden Sie das Ger t bitte mit einer Kontaminationserkl rung zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Wendel...

Page 97: ...genauigkeit 20 0 001mbar 2 v Messwert Reaktionszeit 200 ms Spannungsversorgung 15 30 VDC Elektrischer Anschluss Typ Hirschmann 6polig m nnl verschraubbar Leistungsaufnahme max 1 0 W bei 24 V Versorgun...

Page 98: ...14 VSP63 de 120207 Konformit tserkl rung...

Page 99: ...Thyracont Display Unit 20 3 3 2 Operation With Other Supply And Evaluation Units 21 4 Operation 22 4 1 General 22 4 2 Adjustment 23 4 2 1 Adjustment On Atmosphere and Zero 23 4 2 2 Adjustment On Zero...

Page 100: ...possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversions or modif...

Page 101: ...2m for connection to VD9 display unit W0606002 Measurement cable 6m for connection to VD9 display unit W0606006 Counterplug 6pole 90 bent XB0600002 2 3 Product Description The VSP63 vacuum transducer...

Page 102: ...class A and therefore can cause radio interference in living quarters Improper Use The use for purposes not covered above is regarded as improper in particular the connection to components not allowe...

Page 103: ...again during maintenance work Make vacuum connection via small flange DN16 ISO KF VSP63MV VSP63MVSW or DN16CF Flansch VSP63MVCF If you use the hose barb of model VSP63MVSW be sure to secure the hose...

Page 104: ...F flange connections with elastomer sealings cannot withstand such pressures Process media thus can leak and possibly damage your health 3 3 Electrical Connection 3 3 1 Operation With Thyracont Displa...

Page 105: ...considering EMI demands and according to the pin description shown below Pin Description 1 2 3 6 5 4 Socket Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Type Hirschmann 6pole male Line Terminating Resistor 3...

Page 106: ...y of the unit it is appropriate to allow for stabilization time of 2 5 minutes especially when extreme pressure changes have occurred Accuracy The unit is adjusted ex works in upright position and at...

Page 107: ...ch adjustment point is relevant For zero adjustment actual pressure should be less than 5 0x10 5 mbar alternatively the unit can be adjusted on a variable zero pressure reference see below To achieve...

Page 108: ...ressure underrange By pushing the adj button several times you can now increment this value to maximal 1 0x10 2 mbar 3 according to the pressure voltage relation in chapter 4 1 Then the output again s...

Page 109: ...P63 please send the instrument to us for repair and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or break...

Page 110: ...or 2 from reading Repeatability 20 0 001mbar 2 from reading Response Time 200 ms Power Supply 15 30 VDC Electrical Connection Type Hirschmann 6pin male lockable Power Consumption max 1 0 W with 24 V s...

Page 111: ...27 VSP63 de 120207 Declaration of Conformity...

Page 112: ...VSP63 de 120207...

Page 113: ...VSP63 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...

Page 114: ...Thyracont Anzeigeger te 7 3 3 2 Kundeneigene Spannungsversorgung 8 4 Betrieb 9 4 1 Allgemeines 9 4 2 Nachjustieren 10 4 2 1 Abgleich auf Atmosph re und Nullpunkt 10 4 2 2 Abgleich auf einen bestimmten...

Page 115: ...gliche Reaktionen zwischen Werkstoffen und Prozessmedien Ber cksichtigen Sie m gliche Reaktionen der Prozessmedien infolge der Eigenerw rmung des Produkts Ger t nicht eigenm chtig umbauen oder ver nd...

Page 116: ...zum Anschluss an Anzeigeger t VD9 W0606006 Gegenstecker 6polig gewinkelt XB0600002 2 3 Produktbeschreibung Der Vakuum Messumformer VSP63 dient zur Absolutdruck Messung im Bereich 1000 1 0x10 4 mbar Da...

Page 117: ...rungen f hren Nicht bestimmungsgem e Verwendung Als nicht bestimmungsgem gilt der Einsatz zu Zwecken die von oben genannten abweichen insbesondere der Anschluss an Ger te oder Komponenten die laut ihr...

Page 118: ...leinflansch DN16 ISO KF VSP63MV VSP63MVSW oder DN16CF Flansch VSP63MVCF herstellen Bei Verwendung des Schlauchanschlusses von Modell VSP63MVSW ist die Schlauchverbindung geeignet zu sichern z B mittel...

Page 119: ...ei KF Flanschverbindungen k nnen Elastomer Dichtringe dem Druck nicht mehr standhalten Dies kann zu Gesundheitssch den durch ausstr mende Prozessmedien f hren 3 3 Elektrischer Anschluss 3 3 1 Anschlus...

Page 120: ...EMV gerecht gem untenstehender Pinbelegung herzustellen Pinbelegung 1 2 3 6 5 4 Stecker Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Typ Hirschmann 6pol m nnl Leitungs Abschluss LA 3 0 kOhm auf GND Signalausga...

Page 121: ...insbesondere nach extremen Druckspr ngen angebracht sein eine Stabilisierungszeit von 2min zu beachten Messgenauigkeit Das Ger t ist ab Werk in stehender Position bei einer Versorgungsspannung von 24V...

Page 122: ...m welchen Justierpunkt es sich handelt Beim Nullabgleich sollte der Ist Druck kleiner 5 0x10 5 mbar sein alternativ kann auf einen einstellbaren Referenzwert abgeglichen werden siehe unten Um optimale...

Page 123: ...Nullpunkt Justierwert entspricht Werkseinstellung 1 4V f r ur underrange Durch mehrmaliges Dr cken der Taste adj kann dieser Wert nun gem des Zusammenhangs zwischen Druck und Ausgangssignal aus Abschn...

Page 124: ...nden Sie das Ger t bitte mit einer Kontaminationserkl rung zur Reparatur an uns Das Ger t ist nicht zur kundenseitigen Reparatur vorgesehen Fehlfunktionen des Ger tes die auf Verschmutzung oder Wendel...

Page 125: ...genauigkeit 20 0 001mbar 2 v Messwert Reaktionszeit 200 ms Spannungsversorgung 15 30 VDC Elektrischer Anschluss Typ Hirschmann 6polig m nnl verschraubbar Leistungsaufnahme max 1 0 W bei 24 V Versorgun...

Page 126: ...14 VSP63 de 120207 Konformit tserkl rung...

Page 127: ...Thyracont Display Unit 20 3 3 2 Operation With Other Supply And Evaluation Units 21 4 Operation 22 4 1 General 22 4 2 Adjustment 23 4 2 1 Adjustment On Atmosphere and Zero 23 4 2 2 Adjustment On Zero...

Page 128: ...possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversions or modif...

Page 129: ...2m for connection to VD9 display unit W0606002 Measurement cable 6m for connection to VD9 display unit W0606006 Counterplug 6pole 90 bent XB0600002 2 3 Product Description The VSP63 vacuum transducer...

Page 130: ...class A and therefore can cause radio interference in living quarters Improper Use The use for purposes not covered above is regarded as improper in particular the connection to components not allowe...

Page 131: ...again during maintenance work Make vacuum connection via small flange DN16 ISO KF VSP63MV VSP63MVSW or DN16CF Flansch VSP63MVCF If you use the hose barb of model VSP63MVSW be sure to secure the hose...

Page 132: ...F flange connections with elastomer sealings cannot withstand such pressures Process media thus can leak and possibly damage your health 3 3 Electrical Connection 3 3 1 Operation With Thyracont Displa...

Page 133: ...considering EMI demands and according to the pin description shown below Pin Description 1 2 3 6 5 4 Socket Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Type Hirschmann 6pole male Line Terminating Resistor 3...

Page 134: ...y of the unit it is appropriate to allow for stabilization time of 2 5 minutes especially when extreme pressure changes have occurred Accuracy The unit is adjusted ex works in upright position and at...

Page 135: ...ch adjustment point is relevant For zero adjustment actual pressure should be less than 5 0x10 5 mbar alternatively the unit can be adjusted on a variable zero pressure reference see below To achieve...

Page 136: ...ressure underrange By pushing the adj button several times you can now increment this value to maximal 1 0x10 2 mbar 3 according to the pressure voltage relation in chapter 4 1 Then the output again s...

Page 137: ...P63 please send the instrument to us for repair and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination or break...

Page 138: ...or 2 from reading Repeatability 20 0 001mbar 2 from reading Response Time 200 ms Power Supply 15 30 VDC Electrical Connection Type Hirschmann 6pin male lockable Power Consumption max 1 0 W with 24 V s...

Page 139: ...27 VSP63 de 120207 Declaration of Conformity...

Page 140: ...VSP63 de 120207...

Reviews: