6 Maintenance and Service
Danger of possibly contaminated parts!
Contaminated parts can cause personal injuries. Inform your-
self regarding possible contamination before you start working.
Be sure to follow the relevant instructions and take care of
necessary protective measures.
The unit requires no maintenance. External dirt and soiling can be removed by a
damp cloth.
Should a defect or damage occur on the VD10, please send the instrument for
repair, enclosing a contamination declaration.
The unit is not planned for customer repair!
Messages
Message
Possible Cause
Measures
"Err1"
defective pressure
transducer
send transducer for repair
"notr"
connection to transducer
interrupted
check transducer,
connectors and cables
"ur"
measurement under range
"or"
measurement over range
"off"
channel not in use,
no transducer connected
if appropriate check
transducer, connectors and
cables
"degas"
VSH82 is in degas mode