3.4 Transducer Connection 0-10 V
The instrument must be switched off before any transducers are
connected. Disregarding this instruction may lead to damage of
the instrument.
To this port Thyracont transducers type VSP63MV for absolute pressure with
0 - 10 V signal output can be connected. The VD10 provides voltage supply for
the transducers.
Amphenol
Pin 1:
Pin 2:
Pin 3:
Pin 4:
Pin 5:
Pin 6:
C91E, 6-pole,female
Identification
n.c.
Signal input, 0-10V
GND
24 VDC, max. 8W
AGND
Transducers connected to the analog signal inputs have a fixed assignment to
channel 1 or channel 2.
If an analog transducer connection is used, the corresponding
channel(s) cannot be applied as an address for digital trans-
ducers at the RS485 port!
Examples:
channel 1:
analog, VSP63MV
channel 2, 3, 4 can be assigned to RS485
channel 2:
analog, VSP63MV
channel 1, 3, 4 can be assigned to RS485
channel 1, 2: analog, VSP63MV
channel 3, 4 can be assigned to RS485
Channel 1
Channel 2