Status Model
Power on Settings
The following instrument status values are set at power on:
Status Byte Register
= 0
Service Request Enable Register
✝
= 0
Standard Event Status Register
= 128 (pon bit set)
Standard Event Status Enable Register
✝
= 0
Execution Error Register
= 0
Query Error Register
= 0
Parallel Poll Enable Register
✝
= 0
✝
Registers marked thus are specific to the GPIB section of the instrument and are of limited use
in an RS232 environment.
The instrument will be in local state with the keyboard active.
The instrument parameters at power on are the same as at last switch off with the exception of
RF OUT which is always off.
If for any reason an error is detected at power up in the non-volatile ram a warning will be issued
and all settings will be returned to their default states as for a *RST command.