In order to ensure correct readings during operation, the Photo Diode operating range
must be set before the laser diode is
Input Source Settings
- the source(s) which control the output from the laser unit:
If
Software
is selected, the u nit responds to software commands a nd the
output to the laser is that se t using the SetPowerSetpoint method, or the
‘OUTPUT’ control on the GUI panel.
If
External (SMA)
is selected, the unit responds to the analog signal on the rear
panel EXT IN SMC connector (0 to 10V).
If
Potentiometer
is selected, the output to the laser is the value set using the
top panel OUTPUT potentiometer.
Display Intensity
- determines the brightness of the control panel display.
Enter a value from 0 to 255.
Note
It is possible to select all, or a combination of sources; the sum of the combination is