3.6 TED350 specific commands
TED350 / page 45
3.6.6 Temperature (only AD590 and LM35) (TEMP)
Programming:
Syntax:
"
:TEMP:SET <NR3>
"
Valid Range: Depends on the instrument type.
Default Value: 25 °C
Assumption: The temperature sensor is set to AD590 or LM35. The operating mode
is set to constant temperature / resistance.
Description: Sets the TEC temperature. Unit: [°C].
Reading the set temperature:
Syntax:
"
:TEMP:SET?
"
Response:
[:TEMP:SET <NR3>]
Assumption: The temperature sensor is set to AD590 or LM35.
Description: Queries the programmed temperature. Unit: [°C].
Reading the minimum temperature:
Syntax:
"
:TEMP:MIN?
"
Response:
[:TEMP:MIN <NR3>]
Assumption: The temperature sensor is set to AD590 or LM35.
Description: Queries the minimum allowed temperature. Unit: [°C].
Reading the maximum temperature:
Syntax:
"
:TEMP:MAX?
"
Response:
[:TEMP:MAX <NR3>]
Assumption: The temperature sensor is set to AD590 or LM35.
Description: Queries the maximum allowed temperature. Unit: [°C].
Reading the actual temperature:
Syntax:
"
:TEMP:ACT?
"
Response:
[:TEMP:ACT <NR3>]
Assumption: The temperature sensor is set to AD590 or LM35.
Description: Queries the actual temperature. Unit: [°C].