1.6 Pre-settings
TED350 / page 10
1.6 Pre-settings
To protect the TEC element and the other components of the setup the maximum
allowed TEC current, the maximum temperature as well as an operating temperature
window may be selected.
1.6.1 Setting the limit value of the TEC current
•
Check the maximum allowed TEC current in the datasheet of your TEC
element.
•
Select the display I
TEC LIM
.
•
Change the limit value of the TEC current to the desired value with the
screwdriver-potentiometer besides the LED I
TEC LIM
.
1.6.2 Setting the maximum allowed temperature
To set the maximum allowed temperature (or the minimum allowed resistance):
•
Check the type of temperature sensor in your setup.
•
Select this sensor in the field marked “SENSOR“.
•
Select the display T
SET LIM
.
•
Change the limit value of the temperature (or resistance) to the desired value
with the screwdriver-potentiometer next to the LED T
SET LIM
.
1.6.3 Setting the temperature window
•
Check the type of temperature sensor in your setup.
•
Select this sensor in the field marked “SENSOR“.
•
Select the display T
WIN
.
•
Change the size of the window to the desired value with the screwdriver-
potentiometer next to the LED T
WIN
. The display shows a “plusminus“ – value,
so if you select i.e. 5.000°C the window will reach from 5°C below the set
value to 5°C above the set value.
•
Activate this function with DIP switch 4 (Refer to chapter 2.1.9 on page 22)