6 Appendix
75
6.1.5 ITC4020 Laser Control
ITC4020
Front Panel
Remote Control
Laser Current Control (Constant Current Mode)
Laser Diode Current Range
0 to 20 A
Compliance Voltage
11 V
Setting Resolution
1 mA
320 µA
Measurement Resolution
1 mA
320 µA
Current Accuracy
±(0.1% r 8 mA)
Ripple and Noise, 10 Hz to 10 MHz, rms, typ.
w/o Noise Reduction Filter
with Noise Reduction Filter
< 10 mA
n.a. (no Noise Reduction Filter)
Drift, 24 h (0 -10 Hz, typ., at Constant Ambient
Temperature)
< 1 mA
Temperature Coefficient
LD Current Limit
Setting Range
20 mA to 20 A
Setting Resolution
1 mA
320 µA
Accuracy
±(0.12% r 12 mA)
Power Monitor Input - Photodiode
Photo Current Measurement Ranges
0 to 2 mA / 0 to 20 mA
Photo Current Measurement Resolution,
2 mA Range
20 mA Range
1 µA
10 µA
32 nA
320 nA
Photo Current Accuracy,
2 mA Range
20 mA Range
±(0.08% r0.5 µA)
±(0.08% r5 µA)
Photodiode Reverse BIAS Voltage
0 to 10 V
Photodiode Input Impedance
~0
W
(Virtual Ground)
Power Monitor Input - Thermopile
Voltage Measurement Ranges
0 to 10 mV / 0 to 100mV / 0 to 1V /
0 to 10V
Voltage Measurement Resolution
1 µV / 10 µV / 100 µV /
1mV
0.16 µV / 1.6 µV / 16 µV /
160 µV
Voltage Measurement Accuracy
±(0.1% r 10 µV / + 100 µV / + 1 mV / + 5 mV)
Voltage Input Impedance
1 M
W
Laser Power Control (Constant Power Mode)
Photo Current Setting Resolution,
Control Range 0 to 2 mA
Control Range 0 to 20 mA
1 µA
10 µA
32 nA
320 nA
Thermopile Voltage Setting Resolution,
Control Range 1 µV to 10 mV
Control Range 10 µV to 100 mV
Control Range 100 µV to 1 V
Control Range 1 mV to 10 V
1 µV
10 µV
100 µV
1 mV
0.16 µV
1.6 µV
16 µV
0.16 mV