© 2018 Thorlabs
54
ITC4000 Series
I/O port with external supply
In the figure above, the output drives a relay in sink operation using a higher external voltage
must be set LOW. The maximum sink current is 400 mA.
When driving a relay, connect your external supply voltage also to pin 6 of the digital I/O jack in
order to connect the internal de-spiking clamping diode in parallel to the relay coil. The internal
+12 V supply (which passes a serial diode) is overridden and non-effective in this case.
In
Digital I/O Port
menu the direction of each port can be set to input or output.
In addition the output level of each I/O port can be set, if configured as output. The actual pin
level of each port pin is monitored and displayed in the third line.
For enhanced controlling, please use the remote control interface from an external PC or
Laptop. See the
Thorlabs “Series 4000 Programmers Reference Manual”
on the included “Ser-
ies 4000 Instrumentation CD”.