![THORLABS ITC4000 Series Operation Manual Download Page 36](http://html.mh-extra.com/html/thorlabs/itc4000-series/itc4000-series_operation-manual_1110748036.webp)
© 2018 Thorlabs
34
ITC4000 Series
U
MOD
modulation voltage at "MODULATION IN BNC" jack (R5)
For example; you can add a bipolar (0...±5Vmax) modulation to an internal adjusted 50% cur-
rent / power value or a unipolar modulation (0...10V) if the internal adjusted "DC" value is zero.
The laser current can be monitored at the “ANALOG CONTROL OUT” jack (R6) at the rear
panel.
Note
If the laser current or the laser power reaches the adjusted limits, the status indicators "I LIMIT"
or "P LIMIT" are shown and the green "LD ON" LED is flashing. The modulation will be clipped
to the adjusted limit values.
Attention
The input voltage at the "Modulation Input" BNC jack must not exceed ±10V !