Page 25
22882-D02
Single-Channel Piezo Controller, 75 V to 150 V
condition. If the overtemp condition occurs, disconnect the load and allow 10 minutes
for the unit to cool down.
4.7 Using the Controller with a Force Sensor
The controller unit can also be used to control a force sensor.
1) In the GUI panel, click the ‘Settings button to display the settings panel and select
the Force Sensing tab.
Fig. 4.6 Piezo settings panel
2) Check the
Force Sense Mode
box to select Force Sensing Mode.
3) Enter the calibration factor for the type of force sensor being used. For example,
if set to 1, the GUI digital display shows a detected force of 0 to 1. The default
setting for this parameter is 30, to be compatible with our FSC102 force sensor,
which is specified to read forces up to 30N.
4) The force sensor may display an offset when no contact is apparent. The
Force
Offset
parameter is used to remove any latent offset, such that the sensor only
detects a force when contact is experienced. Enter the required value to remove
any offset from the force sensor output.
In Force Sensor mode, an 'F' is displayed next to the digital display on the GUI panel.
Fig. 4.7 GUI Display - Force Sensor Mode