dwL.U: Setting Dwell Units
Step 4 in programming your controller is defining dwell
units. Dwell units are time units which are used in
“dwELL” segments (amount of time to remain at a specif-
ic temperature ). See “Setting the Segment Type” for an
explanation on how to set a dwell segment.
To set the dwell units:
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2. Press the SCROLL button until display reads,
“dwL.U.”
3. Press the UP or DOWN button to toggle
between seconds, minutes and hours.
CYC.n: Setting the Number of Cycles
Step 5 in programming the controller is to define the
cycle setting. Set the number of times a group of seg-
ments or programs are to be repeated by following the
steps listed below.
To set the cycle setting:
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2. Press the SCROLL button until display
reads,”CYC.n.”
3. Press the UP or DOWN button to select the
number of cycles you want to run or, press the
DOWN button to select “cont.” so the program
will run continuously.
Setting the Segment Type
Step 6 in programming the controller is to define the first
segment type. There are five segment types; “rmP.r,”
“rmP.t,” “dwEll,” “StEP” or “End.” Proceed with the follow-
ing steps according to the type of segment you have
selected.
rmP.r (Ramp)
To ramp linearly at a set rate to a specified temperature:
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
8 S
EGMENT
P
ROGRAMMABLE
M
ODEL W
/OTP
Note
The program ramp rate is designed to
reduce the heatup rate or cooling rate
that the furnace normally exhibits.
When not using this feature, the fur-
nace will operate at its maximum
heating and cooling capability.
25