THIES 4.3811.00.000 Operating Instructions Manual Download Page 14

4.3811.00.000_US-Anemometer-DSP_eng.doc 

14 - 26 

021511/10/06

 

 

After the access key has been changed, the ULTRASONIC transmits a response which contains 
not only the parameter input but also the access mode. 

After parameters have been changed with the key '00001' or '00234', the ULTRASONIC must be 
reset to the locked state with the command 00KY00000. 

Example: 
00KY00001 
USER ACCESS 

 

Response from ULTRASONIC 

!00KY00001 

  Response 

from 

ULTRASONIC 

00AV00005   

 

Change in averaging time 

00KY00000 
WRITE PROTECTED Response from ULTRASONIC 
!00KY00000 

  Response 

from 

ULTRASONIC 

 

 7.1.4  Device ID  

The device ID specifies the address to which the ULTRASONIC is to respond during serial 
communication. The ID is determined on startup via the external signals at PIN 1, PIN 4 and PIN 2 
(ADIO). The device ID lies in the range from 0..7. Every telegram from the ULTRASONIC starts 
with the device ID. Under certain conditions this provides for bus mode, see 7.1.5 

Bus mode. 

On interrogation the device ID 99 is a generic ID. If this ID is used on interrogation, devices transmit 
a response, regardless of the ID selected. 
 

7.1.5 Bus mode 

The concept of ID-based communication allows ULTRASONICs to be operated in a bus system.  

In bus mode there are no restrictions on parameterisability and program upload. A station can also 
be provided with a new program in bus mode for a specific purpose. With an update the program 
update must be performed for every ULTRASONIC. 

7.2 Data 

acquisition 

The main function of the ULTRASONIC firmware is data acquisition and preparation. For data 
acquisition sound impulses are transmitted by the sensors in a clockwise direction and received by 
the sensor opposite. The propagation time measured is a measure of the velocity. A measuring 
cycle is complete when every sensor has performed transmit and receive once. The complete data 
record is then time-stamped and passed on to the next level. After the plausibility check the 
individual components are calculated and written to the averaging buffer (see 7.2.1 

Averaging). 

For a resulting wind velocity < 0.1m/s the wind direction and wind velocity are set to zero. The wind 
direction 0° is reserved for no wind. If the wind direction is equal to zero with WV > 0.1m/s, the 
interface outputs 360°. 
The last valid instantaneous value of the wind velocity is always used as the output value for this 
criterion.  

7.2.1 Averaging 

Given the high data acquisition rate averaging is recommended in most cases. The averaging 
period is set to 1 second with this device version. 
It is a basic rule that only valid values are written to the averaging puffer. The size of the buffer is 
not determined by the number of data records but by the difference in the time stamp between the 
first and last data record. As a result any missing measured values do not influence the averaging 
result. The content level of the averaging buffer is shown in the status value of the ULTRASONIC. 

Summary of Contents for 4.3811.00.000

Page 1: ...sonic Anemometer 2D 4 3811 00 000 Software version V2 00 or higher Status 10 2006 Operating Instructions 021511 10 06 ADOLF THIES GmbH Co KG Hauptstra e 76 37083 G ttingen Germany Box 3536 3541 37025...

Page 2: ...turn values of ULTRASONIC 13 7 1 3 Access mode 13 7 1 4 Device ID 14 7 1 5 Bus mode 14 7 2 Data acquisition 14 7 2 1 Averaging 14 7 3 Serial data output 15 7 3 1 Data enquiry 15 7 3 2 Fixed telegram f...

Page 3: ...23 12 Dimension Drawing 25 13 Accessories available as optional features 25 Fig Fig 1 Connector assembly 8 Table Table 1 Return values with incorrect interpretation of command 13 Table 2 Access key f...

Page 4: ...by means of metal encapsulated ultrasonic transformers These Operating Instructions relate to the device version 4 3811 00 000 They describe possible applications and settings The Ultrasonic Anemomete...

Page 5: ...different wind velocities and directions over a fixed measurement path As the speed of sound greatly depends on the temperature of the air the propagation time of the sound is measured on each of the...

Page 6: ...linear dependence on the absolute humidity level of the air The amount of water vapour in the air proportionately increases the speed of the sound as H2O molecules only have around half the mass of t...

Page 7: ...c anemometer is fitted on the tube or mast socket The arm of the sonic transformer marked red must be aligned to north This allows a bearing to be taken from an object located to the north e g a build...

Page 8: ...ulating tape between litz wire and shield Cable assembly 2 Position shield to rear on cable sheath if cable diameter permits 3 Solder litz wire to female insert position shield in cable clamp 4 Screw...

Page 9: ...l interface 4 Selection device ID Low active Bit 1 Pull up 1mA internal 5 TX RX Y RS 485 RS 422 Serial interface 6 GND Serial interface 7 Supply 20 28 VAC nom 24 VAC No load voltage 32 VAC 8 Supply 20...

Page 10: ...d velocity corresponds to a deviation in the virtual temperature of 1 K at 20 C With a 6 K temperature deviation there is thus a measuring error of approx 1 for the wind velocity Important Mechanical...

Page 11: ...an RS485 interface for serial communication It can be operated either in full or half duplex mode and at different baud rates A standard terminal program is used for communication with the ULTRASONIC...

Page 12: ...a 5 position value The parameter is always right justified i e it thus has to be padded from the left with zeros Example Telegram No 4 is to be interrogated The relevant command is 00TR00004 cr cr sta...

Page 13: ...ls interrogation mode user mode configuration mode Interrogation mode This mode includes commands which do not influence the parameters of the ULTRASONIC They include for example output of the system...

Page 14: ...pose With an update the program update must be performed for every ULTRASONIC 7 2 Data acquisition The main function of the ULTRASONIC firmware is data acquisition and preparation For data acquisition...

Page 15: ...am The time between the last character in the request telegram and the first character in the data telegram is 0 5ms The response to a data enquiry is delayed by 5ms in this device version 7 3 2 Fixed...

Page 16: ...ow nibble 4 bit are EXOR linked and output as an ASCII value 7 5 Status information 7 5 1 Definition of status byte version 1 The status byte contains information about the current system status The i...

Page 17: ...e In this case the error bits are set in the status values and a defined value possibly output at the analog outputs It is a basic rule that the measured values output are always valid and can be inte...

Page 18: ...prox 3sec 7 11 Bootloader Whenever the ULTRASONIC is restarted the device first starts a bootloader The function of the program is to start program upload For this a specific program containing the ne...

Page 19: ...It is possible for the customer to alter the factory setting of the Ultrasonic Anemometer 2D or to adapt it to new requirements Here it should be borne in mind that with a change in settings the orde...

Page 20: ...BH para5 Selects the Bavaria Hesse measuring instrument addresses id BR para5 Select baud rate Baud rate id BX para5 Select baud rate Baud rate extension id BY para5 Scaling of minimum value for anal...

Page 21: ...tware version id TC para5 TC Temperature correction Temperature correction id TE para5 Propagation time delay East sensor Sensor delay EAST sensor id TN para5 Propagation time delay North sensor Senso...

Page 22: ...onse character 2 i 0 9 Identifier 3 i 0 9 Identifier 4 v 0 9 Wind velocity 10 1 m s 5 v 0 9 Wind velocity 10 0 m s 6 v 0 9 Wind velocity 10 1 m s 7 d 0 9 Wind direction 10 2 degree 8 d 0 9 Wind direct...

Page 23: ...K Resolution 0 1 K in telegrams 14 1 to 14 6 Data output digital Interface RS 485 RS 422 Baud rate 1200 2400 4800 9600 19200 38400 57600 115200 230400 460800 921600 adjustable Output Instantaneous val...

Page 24: ...e Program update possible in bus mode Temperature range Operating temp 40 70 C Storage 50 80 C Operating voltage Supply electronics 12 28 V AC DC 10 approx 2 VA Supply heating 24 V AC DC 15 max 28V ty...

Page 25: ...e with coupling socket on transmitter side The other end of the cable is equipped with core identification rings PC Program Meteo Online 9 1700 98 000 For graphical display of measured values on a PC...

Page 26: ...uptstra e 76 37083 G ttingen Germany P O Box 3536 3541 37025 G ttingen Phone 551 79001 0 Fax 551 79001 65 www thiesclima com info thiesclima com Alterations reserved 4 3811 00 000_US Anemometer DSP_en...

Reviews: