Problem/Alarm Screen
Cause
Suggested Actions
My unit gives me a STD WRAP
NOT HOLDING PRESSURE or
DVT WRAP NOT HOLDING
PRESSURE alarm.
Leak in DVT port due to
O-ring damage.
Check for O-ring damage in DVT port.
O-ring damage will cause a leak, not
allowing wrap to hold pressure.
Leak through STD
compression airline
connector O-ring.
Check the O-ring that is on the STD
compression connector fitting going to
the unit.
Internal fault within the
therapy unit.
Check if the power supply LED is green.
Turn the unit off by disconnecting the
power supply. Wait 10 minutes and
reconnect the power supply.
Try an alternate AC outlet.
Contact ThermoTek customer service if
problem persists.
Problem/Alarm Screen
Cause
Suggested Actions
My unit gives me a STD WRAP
NOT INFLATING or DVT WRAP
NOT INFLATING alarm.
Wrap is loose on the
patient.
If the wrap is “ballooned” on the patient,
it has been applied loosely.
Remove wrap and reapply tightly to fit
the patient.
The correct therapy is
not selected.
Make sure you have selected the correct
therapy and that the therapy is active.
Wraps not connected to
unit.
Make sure the wraps are connected to
your unit.
Leaky wrap.
Make sure your wraps are not leaking
air.
Internal fault within the
therapy unit.
Check if the power supply LED is green.
Turn the unit off by disconnecting the
power supply.
Wait 10 minutes and reconnect the
power supply.
Try an alternate AC outlet.
Contact ThermoTek customer service if
problem persists.
58
Summary of Contents for VascuTherm 4
Page 1: ...VascuTherm 4 User Manual...
Page 2: ......
Page 12: ...Observe all warning and caution labels Never remove the caution warning labels 8...
Page 76: ...This page intentionally left blank...
Page 77: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...