• Disconnect the umbilical hose from the unit and connect the therapy wrap directly to the
unit.
• From the front of the unit, remove the (right) wrap connector from the unit. Press the
connector end to purge 4 ounces of fluid from the wrap into a cup.
NOTE:
Do not spill any fluid onto the unit.
• Re-connect the therapy wrap to the unit and visually check the fluid flow for air bubbles.
Let the system run for 1 minute or so to cool the wrap which determines fluid flow. Verify
all air bubbles have been purged from the fluid lines of the wrap.
21
Disconnect umbilical from unit.
Connect therapy wrap directly to
unit.
Remove (right) wrap connector from unit.
Purge 4 ounces of fluid from the wrap.