3.7
CALIBRATION TRIMMING
Calibration trimming allows qualified personnel to adjust the SMX 100’s
calibration by ± 2°C
in 1°C steps.
Note:
A known reference temperature should be used.
1
Press the
key three times to select SET SCREEN 2 (Section
3.3.1).
2
Use the
key to move to the “CALIB (CODE)” window..
3
To enter the CALIBRATION TRIMMING Screen, press and hold the
key for 8 seconds.
4
Use the
keys to move to the channel which requires calibration
trimming, then use the
or
keys to adjust the current
temperature reading.
5
Press any function key to exit Calibration.
22