EXPLANATION OF CONTROLS
The controls are located in a control enclosure to the right hand side of the bath.
MAIN SWITCH:
This is a combination power switch and circuit breaker, when pressed this switch
allows power to be supplied to the controls. Incorporated in the face of the switch is a
“MAINS ON” light.
CAUTION:
The equipment is fitted with a combination
circuit breaker/switch. If the switch continues to trip, it
may indicate an electrical fault exists. Switch off the
power on consult a qualified technician.
TEMPERATURE CONTROL/SENSOR:
This is a Brainchild BTC901 analogue control scaled a 2ºC increments and has been
configured to provide stable control with minimum of overshoot and quick recovery
time.
Sensor type is a RTD Pt100.
Maximum recommended temperature is 60ºC.
HEAT CYCLE INDICATOR:
This output light is illuminated when power is applied to the heating element.
Green: Element off. Red: Element on.
LIGHT SWITCH:
This is a push button, press to turn light on, press again to turn light off. When
activated, the fluorescent light will be illuminated.