CAUTION: 

The range of equipment 

described here is not to be used in 
conjunction with large quantities of 
water. The introduction of water into the 
glass dish recess could result in 
premature failure of components. 

 
 

WARNING: 

Do not use flammable 

solvents or combustible materials as fire 
or explosion may occur. This equipment  
uses components which may be a 
source of ignition.

 

 

 

FUNCTION 

 
To provide an accurate controlled water temperature of a5ºC to +60ºC.

 

 

PRINCIPLE OF OPERATION 

 
The illuminated flotation bath is designed to be light weight and portable. It is ideally 
suited for use in the Histology laboratory.  
The heater body and surrounding area is insulated to minimise heat loss.

 

 
The desired operating temperature is set at the temperature controller. The 
temperature sensor is located within a spring clip and is designed to rest against the 
glass insert (dish). The signal from the temperature sensor is monitored by the 
temperature controller. The resultant calculated pulsed output is directed to a solid 
state relay that switches power to the heating element located below the base of the 
well.  
The warmed base of the well gently and uniformly heats the water within the glass 
dish. The amount of energy is determined by the temperature sensor (spring loaded 
against the glass dish).  
Energy movement is by convection. This limited effect results in an even temperature 
gradient throughout the bath.  
The low wattage fluorescent lamp is activated by the press of a button.  
 

 
 
 

PERFORMANCE 

 
The operating range for this equipment is a5ºC to 60°C with a control 
accuracy of +/- 0.5ºC 
 
 

MAXIMUM TEMPERATURE 

 
For this product is 60

c. 

 
 
 
 

Summary of Contents for TFB-1

Page 1: ...fic Equipment Pty Ltd ABN 80 000 859 129 10 12 Ross Place Wetherill Park NSW 2164 P O Box 6862 Wetherill Park DC NSW 1851 Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 International...

Page 2: ...rts apply when there is a Possibility of personal injury Caution alerts apply when there is a Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT...

Page 3: ...king material check that it is complete Assemble the components and inspect for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage wit...

Page 4: ...temperature is set at the temperature controller The temperature sensor is located within a spring clip and is designed to rest against the glass insert dish The signal from the temperature sensor is...

Page 5: ...y be removed for cleaning The external case and the stainless steel surfaces may be wiped clean using a damp soft cloth Note Stainless steel is under most conditions extremely resistant to corrosion T...

Page 6: ...icate an electrical fault exists Switch off the power on consult a qualified technician TEMPERATURE CONTROL SENSOR This is a Brainchild BTC901 analogue control scaled a 2 C increments and has been con...

Page 7: ...to the well and fill it with water to a maximum depth of 40mm CAUTION Ensure that the ventilation for the rear cooling fan are not obstructed 2 Place the main switch in the ON position reset and immed...

Page 8: ...ng the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the...

Reviews: