background image

P a g e

 | 

 

18-5-10 

TCH-150 & 460 

INTRODUCTION: 

 
Thank you for selecting this product from the large range of products manufactured in Australia by 
Thermoline Scientific. 
 
The materials incorporated in this design have been carefully selected for their durability, technical 
performance, and practicability, the resultant product is one that we are proud of. 
 

THE MANUAL: 

 
Please take the time to read this manual, it contains information that will allow you to understand 
how the equipment operates, what it can and cannot do, how to operate the controls, and how to 
maintain the equipment. 
Ensure that the manual is kept safe but accessible for current and future operators to refer to. 
 

UNPACKING: 

 
Remove the equipment from the packing material, and check that the delivery is complete. Delivery 
should include the following: 

  1 x Temperature and humidity cabinet. 

  Stainless steel open wire type shelves (TCH-150 = 2 shelves, TCH-460 = 4 shelves). 

  1 x Hose connection fitting. 

  1 x Operating manual. 

  1 x Perspex holder. 

Optional extras: These items could be included as part of the packing list if ordered: 

  Water supply bottle (25 litre bottle complete with hose connection fittings). 

  1 x 4 pin plug for retransmission connection. 

  Circular chart recorder (complete with chart paper and chart pens). 

  Data loggers. 

  Extra shelves. 

 
Retain the packing materials until the equipment has been thoroughly tested. 
 
Notify the detail of any damage without delay to your supplier, or Thermoline Scientific. 
 

LOCATION AND SERVICES REQUIRED: 

 
This section deals with the provision and connection of services that will be required for your 
cabinet to function properly, and advice on where to locate the cabinet. 
 

GENERAL: 

 
The equipment is designed to operate under the following ambient conditions: 

  Indoor use only. 

  Ambient temperature range: 10ºC to 35ºC 

  Ambient humidity range: Maximum relative humidity of 85%RH for temperatures up to 25ºC 

  Optimum performance of the chamber is achieved within ambient temperatures of 20ºC to 

25ºC range, and 45%RH to 55%RH. The cabinet can be used outside these ranges but 
performance levels will be affected. Eg: If ambient temperature levels are high (above 
30ºC), the attainable temperature and humidity levels inside the cabinet will be limited.  

 
 
 
 

Summary of Contents for TCH-150

Page 1: ...t Pty Ltd T A Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 Internat...

Page 2: ...pment be aware of the ALERT BOXES that appear throughout these operating instructions INDEX IMPORTANT INFORMATION Page 1 INDEX Page 1 2 PRODUCT WARRANTY Page 3 INTRODUCTION Page 4 THE MANUAL Page 4 UN...

Page 3: ...HANGING THE SET VALUE Pages 13 ADJUSTING THE VENTS Pages 14 15 SENSOR CORRECTION FOR CALIBRATION PURPOSES Page 16 MEASURING EQUIPMENT Page 16 ADJUSTING THE CALIBRATION OFFSET Page 17 MAINTENANCE CLEAN...

Page 4: ...stralia This warranty does not apply to refrigeration compressor s that are not protected by Thermoline Scientific factory installed VFP Voltage Fluctuation Protection unit 4 Limitation of Liability E...

Page 5: ...ese items could be included as part of the packing list if ordered Water supply bottle 25 litre bottle complete with hose connection fittings 1 x 4 pin plug for retransmission connection Circular char...

Page 6: ...d 10 amp power point WARNING Ensure that the electrical outlet is earthed is the correct voltage and capable of carrying the stated current Model TCH 150 is suitable for bench top or floor use whereas...

Page 7: ...HE CABINET DURING SERVICE ASSEMBLY INSTALLATION ASSEMBLY SHELVES Place the shelves at the required level The shelf runners are fixed at 100mm centres PERSPEX HOLDER This holder is provided so that the...

Page 8: ...er to gain access to the drain valve see photo 3 below The drain valve is shown in the closed position Note Prior to leaving the factory the water pressure regulator has been left fully open as it sho...

Page 9: ...of products within the range shown below in the control envelope NOTE Temperature and humidity inside the cabinet can only be controlled at temperatures above the ambient WARNING Do not use flammable...

Page 10: ...justable vents on top of the cabinet an air inlet and an air exhaust The operator can adjust these vents to increase or decrease the amount of ambient air that is mixed with the air inside the cabinet...

Page 11: ...P a g e 10 18 5 10 TCH 150 460 Figure 1 Section side view of cabinet showing air flow EXPLANATION OF CONTROLS Figure 2 Control Panel Layout...

Page 12: ...and quick recovery times Maximum recommended temperature is 80 C Humidity control This is a Eurotherm 3216 microprocessor based instrument with digital indication of set humidity and operating humidi...

Page 13: ...off the fan and heating elements when the door is open 4 Position the mains switch to the ON position reset and immediately both controllers will go through a short initiation period the two displays...

Page 14: ...e alarm will switch off all forms of heating if the actual temperature exceeds the set temperature by 2 0 C This will be indicated by a red flashing indicator and a scrolling message of Over Temperatu...

Page 15: ...gher humidity both vents will be closed for lower humidity the vents will be more open The ambient temperature and humidity will have a large bearing on the temperature and humidity attainable inside...

Page 16: ...P a g e 15 18 5 10 TCH 150 460...

Page 17: ...the operator access level so that it can be used to correct the displayed operating value If adjustments need to be made use the calibration procedure below MEASURING EQUIPMENT Ensure that the equipme...

Page 18: ...e should be carefully considered If in doubt return the value to zero and recalculate the offset value Press the scroll key again to return to the original screen The offset value will have changed to...

Page 19: ...rocess There are several chemicals which will attack the surface of stainless steel plus the lack of oxygen at the surface will cause rusting corrosion and pitting Note Deionised water is particularly...

Page 20: ...lly there is still enough pressure reduction through the valve even when fully open to reduce bounce When mains water is connected to the cabinet As described previously when town water is connected a...

Page 21: ...65 Fan motor Betts 50D503 86A 5 70 Centrifugal blower wheel 5 66 Cooling fan blade 4 30 MRA Bit 02 pressure regulator 4 32 Water solenoid valve L171B13 ZZ23A 1 8 TECHNICAL ASSISTANCE If you require ad...

Page 22: ...the cabinet for connection to an outside monitoring system which is able to accept the 4 20mA signals representing 0 0 to 100 0 C and 0 to 100 RH Maximum load to be applied to the analogue output is 5...

Reviews: