Thermo Electron Corporation
XIII-8
Orion Star Series Meter
Care and Maintenance
Self-Test Instructions
1 . Power up the meter and wait until the SW revision is displayed, then press
.
2 . The display will then fully light and remain lit .
3 . Once you have visual checked that all display segment have lit press
again
and this will turn off the display .
4 . After visually confirming that all the display elements are turned off, press
again the display will show KEY .
5 . Press each key on the meter . If each key isn’t pressed in 5 seconds it will
display “Err” on the middle line & “033” on the bottom line indicating a key
failure . Press
to clear the error and complete the self-test .
6 . If all the keys are functioning then the meter will restart and go to measurement
mode .
7 . If you get the error 034 during the self-test, check that the shorting cap is
installed on the meter .
8 . Reject the meter if any errors are reported during self-test .