Chapter XIII
Care and Maintenance
The Orion Star Series meters combine simple three-button operation with
exceptional long-life performance . Rugged enough for the most challenging
environmental laboratory and field applications, all housings are dustproof and
splash proof . The benchtop meters are rated IP54 . The handheld meters are rated
IP67 and may be immersed for up to one hour with no water incursion;
they even float!
Regular maintenance should include dusting and wiping the unit with a damp cloth .
A mild water-based detergent may be used if necessary .
Spills should be removed immediately using the proper procedure for the type of spill .
Error Codes & Debugging
If an error code is displayed on the screen, the user must press
to clear .
Error
Code
Description
Press
to Clear
Cause
Trouble Shooting
E001
No Model Number
Yes
Memory Corrupted
Contact Customer Service
E002
No HW Calibration
Yes
Memory Corrupted
Contact Customer Service
E003
Bad HW Calibration Yes
Memory Corrupted
Contact Customer Service
E004
RTC Failure
Yes
HW Transient or Failure
Reset Meter, if error
reoccurs contact Customer
Service
E005
User out of range
Yes
HW Transient
or Failure
Re-enter value. Check
specifications for range of
correct values
E033
Keypad Failure
Yes
Hardware problem or
user missed hitting
one of the keys during
the test
Re-run the self test to
insure you have a hardware
problem, then contact
Customer Service
XIII-1
Orion Star Series Meter
Thermo Electron Corporation
Summary of Contents for Orion Star
Page 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Page 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Page 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Page 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...