![Thermo Orion Star User Manual Download Page 82](http://html.mh-extra.com/html/thermo/orion-star/orion-star_user-manual_1104232082.webp)
Thermo Electron Corporation
XII-6
Orion Star Series Meter
5-Star Meter Specifications (Benchtop & Portable)
5 Star pH/ISE/DO/Cond
pH
Range
- 2.000 to 19.999
Resolution
0.1/0.01/0.001
Relative accuracy
± 0.002
ISE*
Concentration Range
0.0 to 19,900.0
Units
M, mg/L, %, ppb, none
Resolution
1, 2, or 3 sig. Digits
Relative Accuracy
± 0.2 mV or 0.05% whichever is greater
mV/RmV
± 1999.9
DO
Concentration
0.000 to 90.00 mg/L
Relative Accuracy
0.2 mg/L
% Saturation
0.0 to 600
Salinity factor
0 to 45 ppt
Barometric pressure
450 - 850 mm Hg
Conductivity
Range
0 - 5000 mS/cm cell constant dependent
See probe specification for the cell constant
Resistivity
0.0001 to 100 Megohms
Salinity
0.01 to 80.0 ppt NaCl
TDS
0.0 to 19.999.0 mg/L
Cell Constant
0.001 to 199.9
Reference temperature
15 °C, 20 °C or 25 °C (default), autoranging
ORP
mV/RmV/Eh Range
± 1999.9
Resolution
± 0.1
Relative accuracy
± 0.2 mV or 0.5%, whichever is greater
* 5-Star pH/DO/Cond meter is also available in a portable model .