![Thermo 8600 Series Operating And Maintenance Manual Download Page 24](http://html.mh-extra.com/html/thermo/8600-series/8600-series_operating-and-maintenance-manual_1104196024.webp)
8600 Series ______________________________________________________________________________________________________________________Maintenance
Preventive Maintenance for 8600 Series Freezers
Refer to Manual Section
Action
Monthly
Yearly
Every
2 Years
--
Verify ambient temperature, <90°F
;
--
* Adjust door handle for firm latching, as needed
;
Figure 1-4 for probe location
Check and clean probe cover, vacuum relief port, gaskets,
;
5.5, 5.6
hinges and door of ice and snow
More frequent cleaning may be
required, depending on use and
environmental conditions.
5.2
Check air filter. Clean or replace as needed
;
1.7, 3.1.a
Check alarm back-up battery.
;
** Replace
--
Check condenser fan motor for unusual motor noise or vibration.
;
2
* Verify and document calibration, at the minimum, annually.
;
5.3
* Clean condenser compartment and wipe off condenser
;
* Qualified service technicians only
** Dispose of properly, according to all state and federal regulations.
5 - 4