cator will light. Press the Silence key and the alarm indicator
will go off. The Low Battery light will stay on until a future
battery test is performed and passed.
d. BUS Battery Test
To test the charge of the BUS battery:
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until BUS BAT TEST is displayed
in the message center.
3. Press Enter to initiate the test.
TESTING BATT will display during the testing period.
Upon completion of the test the message center will display
BBAT GOOD or BBAT FAIL When a test is failed, the audible
alarm will sound, the alarm indicator and the Low Battery indi-
cator will light. Press the Silence key and the alarm indicator
will go off. The Low Battery light will stay on until a future
battery test is performed and passed.
e. Display Temperature
This function, only available on freezers with the
optional sample probe, allows the user to select which tempera-
ture is displayed in the temperature display window. The
options are CONTROL or SAMPLE.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until DISP CONTROL or DISP
SAMPLE is displayed in the message center.
3. Press up/down arrow to toggle between the two display
selections.
4. Press Enter to save.
If control probe is selected, the temperature display
will be on continuously. If sample probe is selected, the temper-
ature display will be preceded with a letter ‘S’.
f. Clear High Stage Alarm
Should a high stage alarm occurred, it may become neces-
sary to the clear the alarm condition after the condition has
been corrected.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until CLR HS ALARM is dis-
played in the message center.
3. Press Enter to clear the alarm.
8600 Series _________________________________________________________________________________Configuration
3 - 1
Section 3 - Configuration
3.1 Configuration Mode
The Configuration Mode is used for testing and custom
setup of the freezer. The configuration functions listed and
described below may not be necessary in all applications, but
are available if needed. See Chart 3-1 for more detail.
a. High Alarm Test
The high alarm test is used to verify the high alarm will
activate, should the freezer temperature equal or exceed the
high alarm set point.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until HI ALRM TEST is displayed
in the message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to increase until
the high alarm set point has been reached. The audible alarm
will sound and the alarm indicator will flash. Press the Silence
key to silence the alarm.
b. Low Alarm Test
The low alarm test is used to verify the low alarm will acti-
vate, should the freezer temperature equal or become less than
the low alarm set point.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until LO ALRM TEST is dis-
played in the message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to decrease until
the low alarm set point has been reached. The audible alarm
will sound and the alarm indicator will flash. Press the Silence
key to silence the alarm.
c. System Battery Test
To test the charge of the freezer battery:
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until SYS BAT TEST is displayed
in the message center.
3. Press Enter to initiate the test.
TESTING BATT will display during the testing period.
Upon completion of the test the message center will display
BATT GOOD or BATT FAIL When a test is failed, the audible
alarm will sound, the alarm indicator and the Low Battery indi-