Thermo Scientific TZ-32 Zonal Rotor
Cleaning of Rotor and Septa — 43
6. 2.
Sterilization of Rotor
Sterilize the rotor appropriately following Table 6-1.
Do not autoclave the rotor or sterilize it in boiling water because its material may deteriorate and its strength may decrease.
Sterilize Method
Condition
Rotor
Usable:
✔
Not usable:
✖
Autoclaving
121 °C (1.0 kg / cm²) for 20 min
✖
Boiling
15-30 min
✖
Ultraviolet rays
200-300 nm
✔
Gas
Ethylene oxide
✔
Formaldehyde
✔
Chemical solution
Ethanol (70%)
✔
Hydrogen peroxide (3%)
✔
Formalin
✔
6. 3.
Inspection of Rotors
Check the rotor as follows after every use.
•
If the optical adapter/disk is unclean, wipe it with a soft cloth but do not damage it.
•
Replacements of packings and O-rings are available.
•
If the rotor appears to be corroded, deformation, cracks or discoloration is observed, do not use the rotor and call
a Thermo Fisher Scientific authorized sales representative. The rotor may be irreparable depending on the degree
of corrosion and wear.
Table 11
Sterilization
methods
Figure 40
Rotor inspection
points