Thermo Scientific TZ-32 Zonal Rotor
General Description — 33
5. 3. 3. Spinning
CAUTION
Never raise the rotor speed to more than 5000 rpm with the rotor not filled with the solution to full extent (less
than 90%). High speed spinning of the rotor not full of the solution will cause the septa to be damaged. The
procedure of spinning the zonal rotor depends on the type of ultracentrifuge used with the rotor. For spinning
the rotor carefully read and follow the instructions given below and in the instruction manual of the centrifuge.
Speed set to 3000 rpm by acceleration High speed spinning
Speed set to 3000 rpm by deceleration
1. Set to ZONAL operation mode from
panel.
2. Set operating factors for high speed
spinning.
3. Press the START switch.
1. Close the door.
2. Press the VACUUM switch.
3. Press the START switch.
1. When the timer counts down to “0”, the
rotor decelerates to 3000 rpm.
2. Press the VACUUM switch.
NOTE
When it is desired to stop the rotor,
set to the normal operation mode
on the panel.
For details on operating from the panel, refer to the instruction manual of the centrifuge.
5. 3. 4. Injection of gradient solution and sample
1. Mount the seal assembly.
Depress the three extrusions of the seal assembly by fitting them into the three relief grooves. Then, the guide at
the end of the seal assembly enters the rotor shaft. Turn the seal assembly clockwise and release it when it touches
the rubber stopper.
➀
➀
Rubber stopper
Table 7
Operation modes
Figure 28
Mount the seal
assembly