Thermo Scientific TZ-32 Zonal Rotor
General Description — 40
»
When the seal assembly is removed due to an injection trouble and the connection is carried out again.
a. When the rotor is full
Inject a heavy solution from “OUTER WALL” side to push the bubbles out of the shaft at ”CENTER” side.
Open the end of the pipe for eliminating bubbles and inject a heavy solution (normally 20 to 40 ml) with an
injector from “OUTER WALL” side until all the bubbles are eliminated from the solution.
b. When the rotor is not full
Because there is no way to eliminate bubbles from the injections, continue injection. If not, abnormal
injection pressure may be applied to the Rulon seal and cause leakage of solution. In addition when
containing bubbles not only the density gradients but also the fractionated patterns may be disturbed in
the extraction process.
2. Do not feed bubbles into the rotor at unloading
When mounting the seal assembly for unloading after centrifugation, eliminate bubbles at “OUTER WALL” side.
Provide the ‘air cushion’, that is an air layer in the pipe for eliminating bubbles to absorb the shock of flow against
the seal assembly.
3. Control the pressure for injecting density gradient solution.
»
The higher the concentration of solution, the slower the injection speed.
»
It is permitted to measure the injection pressure with a pressure gauge to keep the injection pressure at
1 kg/cm² or less.
4. Precautions on mounting the seal assembly
»
Wipe the spinning Rulon seal surface with a tissue paper moistened with water lightly just before mounting
the seal assembly. This prevents the Rulon seal from wearing abnormally and makes the cap removal easier.
»
Keep the seal assembly as horizontal as possible and center it visually. Lower the seal assembly vertically
onto the Rulon seal surface and mount it quickly taking care not to hit the adapter on the Rulon seal surface.
Never incline the seal assembly when mounting. If the seal assembly is mounted in inclined condition, the tip
of the adapter made of Delrin may hit the Rulon surface and the Rulon may be faulty.
5. The flow resistance of extraction pipe must not be large.
»
The diameter of the extraction pipe must not be small and the length must not be long.
»
The flow cell of the UV monitor must not be a micro flow cell