1
Site Preparation
Environmental Requirements
Thermo Scientific
TSQ 8000 Preinstallation Requirements Guide
9
Environmental Requirements
The normal operating environment for the TSQ 8000 must have the following
characteristics:
• Indoor use only
• Altitude up to 2000 meters
• Maximum relative humidity between 5% and 95% up to 31 °C. The maximum relative
humidity decreases linearly to 67% as the temperature climbs to 35 °C.
Use the following guidelines to ensure your site has the proper environmental conditions for
the system:
1. Ensure that your room temperature is 15–35 °C (59–95 °F). The analytical performance
is only confirmed for these temperatures. For best performance, the operating
temperature should be constant. Use
Table 5
to calculate the amount of heat your system
will generate and ensure your air-conditioning system can handle that amount of heat.
Note
The 250V region power cords will work with the 230V circuits.
IMPORTANT
If you are not operating the TSQ 8000, it is safe to leave the instrument
powered on in a room with a temperature between 5 and 40 °C.
Table 5.
Heat output of system components
Instrument
Heat Output (BTU
per Hr)
Heat Output (W)
TSQ 8000, including foreline pump
4300
1260
TRACE 1300 and TRACE 1310
6830
2000
Computer
1
2050
600
Monitor
1
820
240
Optional mech pump
1880
550
AI/AS 1310
325
95
TriPlus RSH
683
2
or 1366
3
200
2
or 400
3
TriPlus 300 Headspace autosampler
4440
1300
Direct Probe Controller
820
240
1
Power requirements vary by manufacturer.
2
With one power module.
3
With two power modules.