5 Manual operation
Thermo Scientific
14
50134830_03.2012
5 Manual operation
INFORMATION
When filling microtiter plates (1), make sure the
wells are not filled to the rim. Otherwise, some
sample splashing might occur.
During the shaking process, the fixating elements (3) hold the microtiter
plate fixed on the shaker platform (4).
TELESHAKE 1536-6
: three movable and three stationary fixating elements
(Figure 6 and Figure 7),
TELESHAKE 1536-8
: four movable and four stationary fixating elements
(Figure 8 and Figure 9).
The movable fixating elements (3) anchor the microtiter plate by using
spring pressure. They can be opened electromechanically in order to insert
or remove the microtiter plate
Using the frequency selection knob (11), you can open and lock the fixating
elements (3):
.
•
Frequency selection knob (11) on position OFF
Fixating elements (3) open (see Figure 6 and Figure 8)
•
Frequency selection knob (11) on shaking frequency
Fixating elements (3) locked (see Figure 7 and Figure 9)
2
If the power supply is defective, the microtiter plate can also be inserted or removed
manually.