Technical Data
Type STP-120
EC Directives
Refer to note on next page
Applicable Standards
Electromagnetic compatibility
Sizes
Diameter of the carrousel
850 mm
Height
500 – 700 mm
Diameter of the turning circle
670 mm
Weight
Including packing
145 kg
Net (fully equipped)
70 kg
Electrical installation
Rated Voltage (±10%)
100-120V ~ / 220-240 V ~
Frequency
50 – 60 Hz
Power consumption
400 VA
Fuses
115 V (2 x T4A)
230 V (2 x T2A)
Battery
Nickel-Cadmium
(12V 600 mA)
IEC Classification 1010/EN61010-1
Degree of protection against electrical
shock.
1
Pollution class
II
Noise
< 70 dB
Capacity
Reagent stations
Number of reagent vessels
10 (9, if 3 are for paraffin.)
Volume per reagent vessel
1.8 L
Baskets for the specimen holders
Number of baskets
1 (option of 2)
Basket capacity
120 specimen holders
Paraffin stations
Number
2 (option of 3)
Volume
1.8 L
Rated Voltage
24 V ~
Rated power per station
100 VA
Temperature range
50 – 70 °C
Hysteresis margin
0.1 – 2 °C
Temperature cut-off
75 °C (± 4 °C)
Thermal protection
Automatic reset