Orion Star A222 and Star A322 Portable Conductivity Meters

7

Nominal Cell Constant Entry

The nominal cell constant value is used during an automatic conductivity calibration and allows the meter to 
determine which Thermo Scientific Orion conductivity standard is being used for the calibration. 

1.  In the measurement mode, press 

2. Press 

 or 

 to highlight 

COND Channel

 and press 

f3 (select)

.

3. Press 

 or 

 to highlight 

Mode and Settings

 and press 

f3 (select)

.

4. Press 

 or 

 to highlight 

Cell K

 and press 

f3 (select)

.

5. Press 

f3 (select)

 to access the numeric entry screen and enter the nominal cell constant of the conductivity cell. 

a. Press 

 or 

 to highlight a number or decimal point, press 

f3 (enter)

 to 

select the highlighted item and repeat until the nominal cell constant value is shown. 

b. Press 

f2 (done)

 to exit the numeric entry screen. 

6. Press 

f1 (back)

 and then press 

 to return to the measurement mode.

Reference Temperature Selection

The reference temperature is the temperature that all conductivity measurements will be reported at by the meter 
if temperature compensation is enabled. 

1.  In the measurement mode, press 

2. Press 

 or 

 to highlight 

COND Channel

 and press 

f3 (select)

.

3. Press 

 or 

 to highlight 

Mode and Settings

 and press 

f3 (select)

.

4. Press 

 or 

 to highlight 

Ref Temp.

 and press 

f3 (select)

.

5. Press 

 or 

 to highlight 

5 °C 

(Star A322 meter only)

, 10 °C 

(Star A322 meter only)

, 15 °C, 20 °C 

or 

25 °C

  

and press 

f3 (select)

.

6. Press 

 to return to the measurement mode.

Temperature Compensation Selection

The temperature compensation can be turned off or set to Linear, nLFn (non-linear ultra pure non-degassed 
water), nLFu (non-linear ultra pure degassed water) or EP (temperature compensation off and warning is displayed 
if conductivity values are outside EP requirements for ultra pure water) and is used to report all conductivity 
measurements at the selected reference temperature.

1.  In the measurement mode, press 

2. Press 

 or 

 to highlight 

COND Channel

 and press 

f3 (select)

.

3. Press 

 or 

 to highlight 

Mode and Settings

 and press 

f3 (select)

.

4. Press 

 or 

 to highlight 

Temp. Comp

. and press 

f3 (select)

.

5. Press 

 or 

 to highlight 

Off, Linear, nLFn, nLFu 

(Star A322 meter only) 

or

 EP 

(Star A322 meter only)

 

and press 

f3 (select)

.

6. Press 

 to return to the measurement mode.

Summary of Contents for Star A222

Page 1: ...top portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and i...

Page 2: ...ion was successfully completed M 100 Indicates a method is in use and the number of the method being used Cond Indicates the type of measurement and determines the type of calibration that will be per...

Page 3: ...lay Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measure...

Page 4: ...the conductivity cell and standard are ready press f3 start 4 Wait for the conductivity value on the meter to stabilize and stop ashing and perform one of the following actions a Press f2 accept to ac...

Page 5: ...and record the conductivity TDS salinity or resistivity and temperature of the sample c Star A322 meter only If the meter is in timed mode the meter will log measurements at the preselected time inte...

Page 6: ...ric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is shown on...

Page 7: ...ted at by the meter if temperature compensation is enabled 1 In the measurement mode press 2 Press or to highlight COND Channel and press f3 select 3 Press or to highlight Mode and Settings and press...