Orion Star A222 and Star A322 Portable Conductivity Meters

4

Conductivity Calibration

One to five conductivity standards can be used for calibration. Always use fresh standards and select 
standards that are near the sample conductivity. Prepare the conductivity cell according to the instructions in 
the conductivity cell use guide. Connect the conductivity cell and any other electrodes to be used to the meter. 
Power on the meter and set the measurement mode to conductivity. 

Note: 

For an automatic calibration, the nominal cell constant of the conductivity cell must be entered in the 

setup menu before the calibration is performed and Thermo Scientific Orion 100 uS/cm, 1413 uS/cm and/or 
12.9 mS/cm conductivity standards must be used.

Automatic and Direct Calibration

1.  In the measurement mode, press 

f1 (cal)

.

2.  Rinse the conductivity cell and any other electrodes in use with distilled water, blot dry with a lint-free 

tissue and place into the standard.

3.  When the conductivity cell and standard are ready, press 

f3 (start)

.

4.  Wait for the conductivity value on the meter to stabilize and stop flashing and perform one of the 

following actions:

a. Press 

f2 (accept)

 to accept the displayed conductivity value.

b. Press 

f3 (edit)

 to access the numeric entry screen and edit the conductivity standard value. 

i. Press 

 

 or 

 to highlight a number or decimal point, press 

f3 (enter)

 to select 

the highlighted item and repeat until the standard value at the measured temperature is shown. 

ii. Press 

f2 (done)

 to exit the numeric entry screen. 

iii. Press 

f2 (accept)

 to accept the entered conductivity value.

5. Press 

f2 (next)

 to proceed to the next standard and repeat steps 2 through 4 or press 

f3 (cal done)

 to 

save and end the calibration. If five standards are used, the calibration will save and end once the fifth 
conductivity standard value is accepted.

6.  The meter will display the calibration summary including the average calculated cell constant. Press 

f1 (meas)

 to export the data to the calibration log or press 

f2 (print)

 to export the data to the calibration 

log and a printer or computer. The meter will automatically proceed to the measurement mode.

Summary of Contents for Star A222

Page 1: ...top portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and i...

Page 2: ...ion was successfully completed M 100 Indicates a method is in use and the number of the method being used Cond Indicates the type of measurement and determines the type of calibration that will be per...

Page 3: ...lay Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measure...

Page 4: ...the conductivity cell and standard are ready press f3 start 4 Wait for the conductivity value on the meter to stabilize and stop ashing and perform one of the following actions a Press f2 accept to ac...

Page 5: ...and record the conductivity TDS salinity or resistivity and temperature of the sample c Star A322 meter only If the meter is in timed mode the meter will log measurements at the preselected time inte...

Page 6: ...ric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is shown on...

Page 7: ...ted at by the meter if temperature compensation is enabled 1 In the measurement mode press 2 Press or to highlight COND Channel and press f3 select 3 Press or to highlight Mode and Settings and press...

Page 8: ...t Conductivity Channel and press f2 select 5 Press or tohighlightConductivity Resistivity TDS SalinityorTemperatureasthecalibrationtype andpressf2 select 6 The meter will display a list of calibration...

Reviews: