Orion Star A222 and Star A322 Portable Conductivity Meters

5

Measurement

Press 

 while taking a measurement in the continuous measurement mode to freeze the display and 

press 

  a second time to unfreeze the display and continue the measurement. Press 

 while taking 

a measurement to manually export the measurement  to the data log, if the data log is enabled in the 
setup menu.

1.  Ensure the measurement mode is set correctly. To set the measurement, press 

 to scroll through the 

conductivity, TDS, salinity and resistivity measurement modes.

2.  Rinse the conductivity cell electrode and any other electrodes in use with distilled water, blot dry with a 

lint-free tissue and place into the sample.

3.  Start the measurement and wait for it to stabilize. 

a.  If the meter is in 

AUTO-READ

 mode (default setting), press 

 to start the measurement. When 

the 

 icon stops flashing, record the conductivity, TDS, salinity or resistivity and temperature of 

the sample. Press 

 again to start a new measurement.

b.  If the meter is in continuous mode, the meter will immediately start taking a measurement and 

update the display whenever the measurement changes. Wait for the display to show 

ready

 and 

record the conductivity, TDS, salinity or resistivity and temperature of the sample.

c. 

Star A322 meter only

 – If the meter is in timed mode, the meter will log measurements at the 

preselected time interval, regardless of the measurement stability. The meter will update the display 
whenever the measurement changes, so the conductivity, TDS, salinity or resistivity and temperature 
of the sample can be recorded when the display shows 

ready

.

4.  Remove the conductivity cell from the sample, rinse with distilled water, blot dry and place into the 

next sample.

5.  Repeat steps 3 and 4 for all samples.

6.  When all samples have been measured, store the conductivity cell according to its user guide.

Read Type Selection

1.  In the measurement mode, press 

.

2. Press 

  or 

 to highlight 

COND Channel 

and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Mode and Settings 

and press 

f3 (select)

.

4. Press 

 

or

 

 to highlight 

Read Type

 and press 

f3 (select)

.

5. Press 

 

or

 

 to highlight 

Auto

Continuous

 or 

Timed 

(Star A322 meter only)

 and press 

f3 (select)

.

 a. 

Star A322 meter only

 – If 

Timed

 is selected and the time interval needs to be changed – highlight 

Timed

;  

 press   to highlight hours (HH), minutes (MM) or seconds (SS); press 

f3 (edit)

 to access the numeric  

  entry screen; use the numeric entry screen to change the values and press 

f1 (back)

 when the time  

  interval is correct. 

6. Press 

 to return to the measurement mode.

Summary of Contents for Star A222

Page 1: ...top portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and i...

Page 2: ...ion was successfully completed M 100 Indicates a method is in use and the number of the method being used Cond Indicates the type of measurement and determines the type of calibration that will be per...

Page 3: ...lay Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measure...

Page 4: ...the conductivity cell and standard are ready press f3 start 4 Wait for the conductivity value on the meter to stabilize and stop ashing and perform one of the following actions a Press f2 accept to ac...

Page 5: ...and record the conductivity TDS salinity or resistivity and temperature of the sample c Star A322 meter only If the meter is in timed mode the meter will log measurements at the preselected time inte...

Page 6: ...ric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is shown on...

Page 7: ...ted at by the meter if temperature compensation is enabled 1 In the measurement mode press 2 Press or to highlight COND Channel and press f3 select 3 Press or to highlight Mode and Settings and press...

Page 8: ...t Conductivity Channel and press f2 select 5 Press or tohighlightConductivity Resistivity TDS SalinityorTemperatureasthecalibrationtype andpressf2 select 6 The meter will display a list of calibration...

Reviews: