Decontamination
Given the nature of samples processed in a rotor, biological or radioactive contamination is possible.
For biological contamination of rotors, a 2
% glutaraldehyde solution, ethylene oxide or ultraviolet
radiation are the recommended methods of sterilization, While for a rotor that may be contaminated
by a radioactive sample, use a solution of equal parts of 70
% ethanol, 10
% SDS and water. In
addition:
•
Do not use chlorine bleach on aluminum rotors.
•
When autoclaving, rotor components should be separated.
•
If sterilization is not necessary, a 70
% solution of ethanol can be used.
•
Most commercially available detergents for radioisotopic contamination are not compatible with
aluminum or anodized coatings and shall not be used.
•
Rinse with ethanol, followed by water and dry with a soft cloth.
•
Do not immerse Thermo Scientific Fiberlite rotors; spin rotor to remove liquid.
•
Fiberlite composite rotors are not compatible with ethylene oxide.