Configuration Considerations when Using Sarasota CM515
General Configuration (Metric) for SG Measurement
Thermo Fisher Scientific
Sarasota SG901 User Guide
F-9
Table F–7.
Input menu items
Input
Setting
Units
Comments
Display
Parameter Name
Pulse Input 1
Assignment
Density Line
Fixed default
Pulse Input 1 Signal
Type
PULSE
Density pulse
Pulse Input 1 Signal
Debounce
DISABLE
Pulse Input 1
Frequency Cutoff
100
Pulse Input 1
Filtering
0
Analog Input 1
Assignment
Temperature
Fixed default
Analog Input 1
Signal Type
PT100
Analog Signal 1
Default Type
50
Deg C
Assume heated SG system and
set to heater set temperature.
Analog Input 1
Minimum Point
0
Deg C
Analog Input 1
Maximum Point
50
Deg C
Analog Input 2
Assignment
Pressure
Fixed default
Analog Input 2
Pressure Sensor
Type
ABSOL (Absolute)
Analog Input 2
Signal Type
4–20 mA
Analog Input 2
Default
3.8
bar A
Set to the regulator pressure
setting.
Analog Input 2
Minimum Point
0
bar A
Analog Input 2
Maximum Point
4
bar A
Values on Exception
Default
Summary of Contents for Sarasota SG901
Page 2: ......
Page 3: ...Sarasota SG901 Specific Gravity Analyzer User Guide P N HB SG901 Revision E...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 12: ...This page intentionally left blank...
Page 14: ...This page intentionally left blank...
Page 18: ...This page intentionally left blank...
Page 24: ...This page intentionally left blank...
Page 30: ...This page intentionally left blank...
Page 38: ...This page intentionally left blank...
Page 48: ...This page intentionally left blank...
Page 52: ...This page intentionally left blank...
Page 74: ...This page intentionally left blank...
Page 76: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...