Validation
Validation with an Injection Sample
4-2
Sarasota SG901 User Guide
Thermo Fisher Scientific
Figure 4–1.
System validation flow diagram
Follow these steps to validate the system using an injection sample.
1.
Close valves 1 and 2, and open valve 4.
2.
Invert the validation gas sample cylinder several times to ensure a good
mix.
3.
Connect the gas cylinder to the validation gas input via a pressure
regulator set to the same pressure as the Sarasota SG901 regulator.
4.
Allow the sample to flow at approximately 5 L/min until the reading
from the density converter is stable. Note the SG or MW reading, and
compare it to the certified SG or MW. The error between the system
and the certified value should be within the sum of the uncertainties of
the validation gas and the Sarasota SG901. If the error is unacceptable,
go to
Chapter 7
for troubleshooting.
5.
After validation, close the validation cylinder valve and valve 3. Open
valves 2 and 1.
Validation with
an Injection
Sample
Summary of Contents for Sarasota SG901
Page 2: ......
Page 3: ...Sarasota SG901 Specific Gravity Analyzer User Guide P N HB SG901 Revision E...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 12: ...This page intentionally left blank...
Page 14: ...This page intentionally left blank...
Page 18: ...This page intentionally left blank...
Page 24: ...This page intentionally left blank...
Page 30: ...This page intentionally left blank...
Page 38: ...This page intentionally left blank...
Page 48: ...This page intentionally left blank...
Page 52: ...This page intentionally left blank...
Page 74: ...This page intentionally left blank...
Page 76: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...