Calibration
Adjustment
5-4
Sarasota SG901 User Guide
Thermo Fisher Scientific
Gas
Pressure 25°C
30°C
35°C
40°C
45°C
2 bar A
1.0012
1.0012
1.0012
1.0012
1.0012
H2
3 bar A
1.0018
1.0018
1.0018
1.0018
1.0018
4 bar A
1.0024
1.0024
1.0024
1.0024
1.0024
2 bar A
0.9898
0.9904
0.9909
0.9913
0.9918
CO2
3 bar A
0.9847
0.9855
0.9863
0.9870
0.9877
4 bar A
0.9796
0.9807
0.9817
0.9827
0.9836
2 bar A
0.9955
0.9997
0.9998
0.9998
0.9999
N2
3 bar A
0.9933
0.9996
0.9997
0.9998
0.9999
4 bar A
0.9911
0.9994
0.9996
0.9997
0.9998
Table 5–2.
Gas compressibility at reference conditions (0°C and 1.01325 bar A)
C1
C2
C3
H2
CO2
N2
0.9976
0.9986
0.9588
1.0006
0.9937
0.9995
Summary of Contents for Sarasota SG901
Page 2: ......
Page 3: ...Sarasota SG901 Specific Gravity Analyzer User Guide P N HB SG901 Revision E...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 12: ...This page intentionally left blank...
Page 14: ...This page intentionally left blank...
Page 18: ...This page intentionally left blank...
Page 24: ...This page intentionally left blank...
Page 30: ...This page intentionally left blank...
Page 38: ...This page intentionally left blank...
Page 48: ...This page intentionally left blank...
Page 52: ...This page intentionally left blank...
Page 74: ...This page intentionally left blank...
Page 76: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...