4-2
Revco CxF
Thermo Scientific
Section 4
Alarms
This condition is created when the high stage compressor and fans run for
30 minutes and are not capable of cooling the interstage heat exchanger to
the proper temperature. Under this condition, the high stage compressor
and fans will turn off after 30 minutes and an audible and visual alarm will
occur along with the "HS SYST FAIL" message in the LED message
center.
When multiple alarm conditions occur, active messages are displayed in the
message center one at a time, updating at 5 second intervals. Pressing
Silence during multiple alarms causes all active alarms to be silenced and to
ring back in 15 minutes.
An internal communication failure has occurred with the micro board.
During this alarm, the compressor(s) attempt to run continuously.
However, with this type of failure, freezer operation becomes
undependable.
Communication between the micro board and the display board has been
lost. Under this condition, the visual alarm flashes along with dashes in the
temperature display (----). Contact Technical Services.
High Stage System Failure
Alarm (-86C units only)
Multiple Alarms
Micro Board Failure
Alarm
Lost Communication
Summary of Contents for Revco ULT-1390-10-A
Page 54: ...Revco CxF 8 1 Thermo Scientific Section 8 Spare Parts...
Page 73: ...Revco CxF 9 1 Thermo Scientific Section 9 Refrigeration Schematics...
Page 74: ...9 2 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 75: ...Revco CxF 9 3 Thermo Scientific Section 9 Refrigeration Schematics...
Page 76: ...9 4 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 77: ...Revco CxF 9 5 Thermo Scientific Section 9 Refrigeration Schematics...
Page 78: ...9 6 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 79: ...Revco CxF 10 1 Thermo Scientific Section 10 Electrical Schematics...
Page 80: ...10 2 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 81: ...Revco CxF 10 3 Thermo Scientific Section 10 Electrical Schematics...
Page 82: ...10 4 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 83: ...Revco CxF 10 5 Thermo Scientific Section 10 Electrical Schematics...
Page 84: ...10 6 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 85: ...Revco CxF 10 7 Thermo Scientific Section 10 Electrical Schematics...
Page 86: ...10 8 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 87: ...Revco CxF 10 9 Thermo Scientific Section 10 Electrical Schematics...
Page 88: ...10 10 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 89: ...Revco CxF 10 11 Thermo Scientific Section 10 Electrical Schematics...
Page 90: ...10 12 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 91: ...Revco CxF 10 13 Thermo Scientific Section 10 Electrical Schematics...
Page 92: ...10 14 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 93: ...Revco CxF 10 15 Thermo Scientific Section 10 Electrical Schematics...
Page 94: ...10 16 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 95: ...Revco CxF 10 17 Thermo Scientific Section 10 Electrical Schematics...
Page 96: ...10 18 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 97: ...Revco CxF 10 19 Thermo Scientific Section 10 Electrical Schematics...
Page 98: ...10 20 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 99: ...Revco CxF 10 21 Thermo Scientific Section 10 Electrical Schematics...
Page 100: ...10 22 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 101: ...Revco CxF 10 23 Thermo Scientific Section 10 Electrical Schematics...
Page 102: ...10 24 Revco CxF Thermo Scientific Section 10 Electrical Schematics...