4-4
Revco CxF
Thermo Scientific
Section 4
Alarms
Error
(cont.)
High End Message
Notes
Er02
“CNT PRB FLT”
Name:
Control (Cabinet) Sensor Failure
Description:
This condition indicates that the control sensor has failed to produce a valid reading for >12
consecutive reads (~60 seconds).
Response:
The unit will stage both compressors on (if necessary) and the unit will attempt to head to bot-
tom out. If the sensor recovers, the system will begin to operate normally and respond to the temperature
feedback. The remote alarm contacts will become active regardless of the key position for this mode of fail-
ure. ‘Er02’ will be added to the main display queue and the last valid cabinet temperature value will not be
displayed
Er03
“HSHX PRB FLT”
-86C units only
Name:
Heat Exchange Sensor Failure
Description:
This condition indicates that the heat exchange sensor has failed to produce a valid reading
for >12 consecutive reads (~60 seconds).
Response:
The display will show “Er03” only when the button sequence to read the heat exchange sensor
is depressed.
Er05
N/A
Name:
Display Firmware Integrity Failure
Description:
The display firmware has failed to pass its CRC CCITT checksum integrity test.
Response:
The display performs this check at startup and the display board will fail to startup with out any
error indication if it does not pass this at power on.
Er06
N/A
Name:
Micro Firmware Integrity Failure
Description:
The micro firmware has failed to pass its CRC CCITT checksum integrity test.
Response:
This is checked at power on reset and the “Er06” will be displayed for ~10 seconds at startup if
this condition exists.
Er07
“MICRO FAIL”
Name:
Micro Fail - CS5521 SPI Failure / UISR Failure
Description:
This condition indicates a micro board failure due to either the SPI bus is unable to communi-
cate with the ADC device or a UISR event caused the microcontroller to be in an unstable state.
Response:
The unit will try to recover from this fault three times by a hardware reset of the micro board. In
the event that the system couldn’t rectify the issue, the following sequence of events will occur:
1. Remote alarm contacts will become active.
2. Buzzer will annunciate audibly and will have a ringback of 15 minutes.
3. “Seven segment” display will show “Er07”.
4. The -86C system will have 10 minute staging between the high stage compressor and the low stage
compressor activation (-40C units will activate the compressor).
5. The system will go to bottom out temperatures.
Summary of Contents for Revco ULT-1390-10-A
Page 54: ...Revco CxF 8 1 Thermo Scientific Section 8 Spare Parts...
Page 73: ...Revco CxF 9 1 Thermo Scientific Section 9 Refrigeration Schematics...
Page 74: ...9 2 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 75: ...Revco CxF 9 3 Thermo Scientific Section 9 Refrigeration Schematics...
Page 76: ...9 4 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 77: ...Revco CxF 9 5 Thermo Scientific Section 9 Refrigeration Schematics...
Page 78: ...9 6 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 79: ...Revco CxF 10 1 Thermo Scientific Section 10 Electrical Schematics...
Page 80: ...10 2 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 81: ...Revco CxF 10 3 Thermo Scientific Section 10 Electrical Schematics...
Page 82: ...10 4 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 83: ...Revco CxF 10 5 Thermo Scientific Section 10 Electrical Schematics...
Page 84: ...10 6 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 85: ...Revco CxF 10 7 Thermo Scientific Section 10 Electrical Schematics...
Page 86: ...10 8 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 87: ...Revco CxF 10 9 Thermo Scientific Section 10 Electrical Schematics...
Page 88: ...10 10 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 89: ...Revco CxF 10 11 Thermo Scientific Section 10 Electrical Schematics...
Page 90: ...10 12 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 91: ...Revco CxF 10 13 Thermo Scientific Section 10 Electrical Schematics...
Page 92: ...10 14 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 93: ...Revco CxF 10 15 Thermo Scientific Section 10 Electrical Schematics...
Page 94: ...10 16 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 95: ...Revco CxF 10 17 Thermo Scientific Section 10 Electrical Schematics...
Page 96: ...10 18 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 97: ...Revco CxF 10 19 Thermo Scientific Section 10 Electrical Schematics...
Page 98: ...10 20 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 99: ...Revco CxF 10 21 Thermo Scientific Section 10 Electrical Schematics...
Page 100: ...10 22 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 101: ...Revco CxF 10 23 Thermo Scientific Section 10 Electrical Schematics...
Page 102: ...10 24 Revco CxF Thermo Scientific Section 10 Electrical Schematics...