Configuration Mode is used for testing and custom setup of the freezer.
The configuration functions listed and described below may not be
necessary in all applications, but are available if needed. See Chart 3-1 for
more detail.
The high alarm test is used to verify that the high alarm will activate,
should the freezer temperature equal or exceed the high alarm set point.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until HI ALRM TEST is displayed in the
message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to increase until the high alarm
set point has been reached. The audible alarm will sound and the alarm
indicator will flash. Press the Silence key to mute the alarm.
The low alarm test is used to verify the that low alarm will activate, should
the freezer temperature equal or become less than the low alarm set point.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until LO ALRM TEST is displayed in the
message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to decrease until the low alarm
set point has been reached. The audible alarm will sound and the alarm
indicator will flash. Press the Silence key to mute the alarm.
High Alarm Test
Low Alarm Test
Summary of Contents for Revco ULT-1390-10-A
Page 54: ...Revco CxF 8 1 Thermo Scientific Section 8 Spare Parts...
Page 73: ...Revco CxF 9 1 Thermo Scientific Section 9 Refrigeration Schematics...
Page 74: ...9 2 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 75: ...Revco CxF 9 3 Thermo Scientific Section 9 Refrigeration Schematics...
Page 76: ...9 4 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 77: ...Revco CxF 9 5 Thermo Scientific Section 9 Refrigeration Schematics...
Page 78: ...9 6 Revco CxF Thermo Scientific Section 9 Refrigeration Schematics...
Page 79: ...Revco CxF 10 1 Thermo Scientific Section 10 Electrical Schematics...
Page 80: ...10 2 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 81: ...Revco CxF 10 3 Thermo Scientific Section 10 Electrical Schematics...
Page 82: ...10 4 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 83: ...Revco CxF 10 5 Thermo Scientific Section 10 Electrical Schematics...
Page 84: ...10 6 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 85: ...Revco CxF 10 7 Thermo Scientific Section 10 Electrical Schematics...
Page 86: ...10 8 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 87: ...Revco CxF 10 9 Thermo Scientific Section 10 Electrical Schematics...
Page 88: ...10 10 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 89: ...Revco CxF 10 11 Thermo Scientific Section 10 Electrical Schematics...
Page 90: ...10 12 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 91: ...Revco CxF 10 13 Thermo Scientific Section 10 Electrical Schematics...
Page 92: ...10 14 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 93: ...Revco CxF 10 15 Thermo Scientific Section 10 Electrical Schematics...
Page 94: ...10 16 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 95: ...Revco CxF 10 17 Thermo Scientific Section 10 Electrical Schematics...
Page 96: ...10 18 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 97: ...Revco CxF 10 19 Thermo Scientific Section 10 Electrical Schematics...
Page 98: ...10 20 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 99: ...Revco CxF 10 21 Thermo Scientific Section 10 Electrical Schematics...
Page 100: ...10 22 Revco CxF Thermo Scientific Section 10 Electrical Schematics...
Page 101: ...Revco CxF 10 23 Thermo Scientific Section 10 Electrical Schematics...
Page 102: ...10 24 Revco CxF Thermo Scientific Section 10 Electrical Schematics...