RadEye PRD-ER
DB-066 E
4-13
Iw/Tr/Ff 06.07.2015
“Alk.” (Alkaline) has a threshold of 2.1V. Type “NiMH” has a
threshold of 2.35V. See also chapter 5.3.
Battery type “Alkaline”
Battery type “NiMH”
4.16
Lu-Test
A “Test Adapter RadEye PRD” part number 42506/7071 is
needed to perform this test. The procedure of the Lu-test is
described in the manual of the test adapter. Purpose of this test
is the verification of the precise performance of the detector
amplification. Small variations are corrected automatically,
while larger deviations from the original settings will result in
a LCD-message, that the instrument should be inspected for a
possible detector problem which may influence the accuracy
of the dose rate readings.
Start the Lu-Test after learning phase via submenu “Settings” -
> “Lu-Test” without the test adapter.
Date and time of the last
successful Lu-Test