3-6
DB-066 E
RadEye PRD-ER
Iw/Tr/Ff 06.07.2015
- 3
rd
click:
the mean value and max value of the measured
count rate (cps), together with the time
indicated since the last reset.
After 10 seconds or after 4
th
click standard display with count
rate indication is activated again.
1
st
click
2
nd
click
3
rd
click
3.2.1.2
Count Rate Mode with Dose Rate Display
In this mode the measurement value of the dose rate is displayed
together with the bar graph of the cps value. The count rate alarm
thresholds are active. That is the alarming has a quick alarm re-
sponse, while the calculating of the dose rate for low count rates
takes place over a longer time. (see chapter 3.2.2).
This mode is for rapid finding of radioactive sources while
displaying true dose rate values. A press on the hot key (arrow
up) results directly to the display of the accumulated dose
respectively the mean- and max value.
This mode can be activated or deactivated alternatively to the
Count Rate Mode (3.2.1) via PC-Software only.