RadEye PRD-ER
DB-066 E
3-11
Iw/Tr/Ff 06.07.2015
3.5
Setting alarm thresholds
The menu options
Alarm Cnt. Rate
,
Alarm Dose
Rate
,
Alarm Dose
and
Alarm level
allow the alarm thresholds
to be modified. For this setting, the user has 255 seconds time.
Changing the value is effected by pressing the “
◄
” (Change)
button if the corresponding “Alarm” is selected:
To increment the number, press the “
▲
/
▼
” arrow keys. To go
on to the next digit or to quit the edit mode, menu use “
◄
/
►
”
arrow keys.
Once the last number has been set, quit the editing mode by
pressing the “Exit”-key. Then, the value set is saved and the
unit returns to the basic display.
Example:
The dose alarm threshold must be changed.
Press “Menu” key, and “
▲
/
▼
”
arrow keys until
Alarm Dose
is selected
Then enter change menu by press-
ing “Change” key.
Edit value by pressing “
▲
/
▼
” keys
or select digit by “
◄
/
►
” keys.
Pre unit “
µ
”, “m” may be chosen as
well.