2-2
DB-066 E
RadEye PRD-ER
Iw/Tr/Ff 06.07.2015
2.2
Inserting the battery
The two AAA-Alkaline cells as delivered with the instrument
allow up to 600 h of normal operation.
AAA size rechargeable batteries can be used as well.
•
Switch off the measuring instrument.
•
Remove rubber boot.
•
Open the cover of the battery compartment.
Use of a coin is recommended.
•
Exchange the batteries according to the shown polarity.
Always replace both batteries with new ones and the same
type.
•
Close the compartment cover hooks first, care for the rub-
ber seal being in its groove.
•
Switch on the unit again. (see chapter 2.4)
The instrument continues working in the operating mode set
last (see chapter 3.2).
The measured values in the history memory remain stored.