11. Technical data
11.1
RadEye GF /GF10
Radiation type:
Gamma and X-ray radiation
Measured quantity:
Ambient equivalent dose rate:
RadEye GF:
x [R/h] and dose:
H
x[R]
RadEye GF-10:
*(10) [Sv/h] and dose:
H
*(10)[Sv]
Please Note:
If the native unit (see above) is changed
by menu, the displayed measurement
values are converted. This is not a con-
version with energy correction!
Measuring range:
Dose Rate:
3 Sv/h, 300 R/h, 300 rem/h or 2.6 Gy/h
Dose:
10 Sv
Overload display:
more than 3 Sv/h or 300 R/h
overload indication up to 10 Sv/h
Linearity error:
max.
±
10 % in the measuring range