4-10
DB-083 E
RadEye GF-Series
Le/Ff 20.07.2015
4.14
Source test
Source test is started with menu item “Source test” in submenu
“Settings”.
Source test starts with the parameters previously set with
radeye.exe. Furthermore it is necessary to set the actual date
and time.
Start the source test without source. The RadEye check the
background value. If the background is lower than 0.15 µSv/h
the user is asked to place the source.
After placing the source, source test is started automatically.
Date and time of last success-
ful source test
Low range test
High range test
Terminate
source test with
“
►
” key
Start source test
by placing the
source